Calculative technique to correct for the change in linear accelerator beam energy at off-axis points
- 1 March 1980
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 7 (2) , 147-150
- https://doi.org/10.1118/1.594663
Abstract
The change in energy of linear accelerator X-ray beams from the central ray to off-axis points causes errors in the dose calculated by conventional techniques for large, irregularly shaped fields. A modification of conventional radiotherapy calculative methods to correct for the change in beam energy is presented. Results of measurements in irregular fields on a Clinac-4 which verify the validity of the calculative method are reported. A discussion of the clinical significance will point out errors of 3-4% in conventional dose calculations.Funding Information
- National Cancer Institute (CA 10953)
- DHEW
This publication has 1 reference indexed in Scilit:
- Off-axis beam quality change in linear accelerator x-ray beamsMedical Physics, 1980