Total Knee Arthroplasty With 4.4 mm of Tibial Polyethylene
- 1 July 2001
- journal article
- Published by Wolters Kluwer Health in Clinical Orthopaedics and Related Research
- Vol. 388 (388) , 112-117
- https://doi.org/10.1097/00003086-200107000-00017
Abstract
A consecutive series of 387, one-piece, 8-mm tibial components were implanted in 313 patients. All tibial prostheses were manufactured with 4.4 mm of polyethylene directly molded to a 3.6 mm cobalt chrome metal baseplate. The average age of the patients at surgery was 70.6 years. From this group, a subset of 116 patients underwent bilateral total knee arthroplasties with an 8-mm tibial component on one side and at least a 10-mm thick tibial component (at least 6.4 mm of polyethylene) on the other side. The followup averaged 10.7 years. Postoperative knee and pain scores average 81.4 points and 47.2 points, respectively. No radiographic polyethylene wear or osteolytic lesions were identified. Tibial radiolucencies were observed in four knees. There were nine failures (2.3%). Three knees were revised for infection. Five knees were revised for metallosis secondary to failure of metal-backed patellas. No loose tibial components were identified. Defining failure as revision for any reason or loosening of any component, Kaplan-Meier survival rates at 5, 10, and 15 years were 98.7%, 95.4%, and 94.3%, respectively. There was no significant difference in clinical scores or survival rates between the knee components with thicker (at least 6.4 mm) or thinner (4.4 mm) polyethylene.Keywords
This publication has 13 references indexed in Scilit:
- Dissemination of Wear Particles to the Liver, Spleen, and Abdominal Lymph Nodes of Patients with Hip or Knee Replacement*Journal of Bone and Joint Surgery, 2000
- Plasticity-induced damage layer is a precursor to wear in radiation-cross-linked UHMWPE acetabular components for total hip replacementThe Journal of Arthroplasty, 1999
- Effects of machining on tribological behavior of ultra high molecular weight polyethylene (UHMWPE) under dry reciprocating slidingWear, 1999
- Current Concepts Review - Wear in Total Hip and Knee Replacements*Journal of Bone and Joint Surgery, 1999
- Modular Tibial Insert MicromotionPublished by Wolters Kluwer Health ,1998
- Tibial Insert Undersurface as a Contributing Source of Polyethylene Wear DebrisClinical Orthopaedics and Related Research, 1997
- The posterior stabilized total knee prosthesis. Assessment of polyethylene damage and osteolysis after a ten-year-minimum follow-up.Journal of Bone and Joint Surgery, 1995
- The Knee Society Total Knee Arthroplasty Roentgenographic Evaluation and Scoring SystemClinical Orthopaedics and Related Research, 1989
- The Problem of Surface Damage in Polyethylene Total Knee ComponentsClinical Orthopaedics and Related Research, 1986
- Nonparametric Estimation from Incomplete ObservationsJournal of the American Statistical Association, 1958