We report high-precision isotopic carbon dioxide measurements, made before and after ion source modification to gas isotope ratio mass spectrometry (IRMS) instruments. Measurement protocols were designed to explore the effects of ion source material substitution, source conductance, inlet pressure, electron emission, acceleration potential, and inlet changeover equilibration time. After modification of the IRMS instruments at the National Institute of Standards and Technology (NIST) and the Max-Planck-Institute for Chemistry (MPI-Mainz), immediate changes were observed. At NIST, measurements were no longer sensitive to inlet equilibration times greater than 15 s, and different settings of ion source conductance resulted in δ13C shifts of about 0.04‰ per 10‰ measurement difference between sample and reference, a five-fold improvement. No significant changes in machine performance were observed after a month of use. After a year, performance had degraded slightly, but was controlled by ion source cleaning and the use of low-energy ion acceleration to minimize sputtering. At MPI-Mainz, results were very similar. We report cross-contamination coefficients measured since 1996, and discuss the role of adsorption, ion implantation, and sputtering on cross contamination in mass spectrometry systems. We recommend that users of high-precision IRMS instruments test for and minimize the effects described. Published in 2003 by John Wiley & Sons, Ltd.Keywords
This publication has 16 references indexed in Scilit:
- Isotopic metrology of carbon dioxide. I. Interlaboratory comparison and empirical modeling of inlet equilibration time, inlet pressure, and ion source conductanceRapid Communications in Mass Spectrometry, 2003
- The growth and modification of materials via ion–surface processingSurface Science, 2002
- web-based interactive data processing: application to stable isotope metrologyAnalytical and Bioanalytical Chemistry, 2001
- Cross contamination in dual inlet isotope ratio mass spectrometersInternational Journal of Mass Spectrometry, 2000
- Trends in sputteringProgress in Surface Science, 2000
- Preparation, Characterization, and Value Assignment of Carbon Dioxide Isotopic Reference Materials: RMs 8562, 8563, and 8564Analytical Chemistry, 1999
- Isotopic fractionation in low‐energy ion implantationJournal of Geophysical Research, 1997
- Sputtering yield measurementsPublished by Springer Nature ,1981
- Proportional variations of geochemical δ 18O scales —an interlaboratory comparisonGeochimica et Cosmochimica Acta, 1978
- The measuring procedure and corrections for the high-precision mass-spectrometric analysis of isotopic abundance ratios, especially referring to carbon, oxygen and nitrogenInternational Journal of Mass Spectrometry and Ion Physics, 1973