Actual flip‐angle imaging in the pulsed steady state: A method for rapid three‐dimensional mapping of the transmitted radiofrequency field
Top Cited Papers
Open Access
- 26 December 2006
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 57 (1) , 192-200
- https://doi.org/10.1002/mrm.21120
Abstract
A new method has been developed for fast image‐based measurements of the transmitted radiofrequency (RF) field. The method employs an actual flip‐angle imaging (AFI) pulse sequence that consists of two identical RF pulses followed by two delays of different duration (TR1 < TR2). After each pulse, a gradient‐echo (GRE) signal is acquired. It has been shown theoretically and experimentally that if delays TR1 and TR2 are sufficiently short and the transverse magnetization is completely spoiled, the ratio r = S2/S1 of signal intensities S1 and S2, acquired at the beginning of the time intervals TR1 and TR2, depends on the flip angle (FA) of applied pulses as r = (1 + n * cos(FA))/(n + cos(FA)), where n = TR2/TR1. The method allows fast 3D implementation and provides accurate B1 measurements that are highly insensitive to T1. The unique feature of the AFI method is that it uses a pulsed steady‐state signal acquisition. This overcomes the limitation of previous methods that required long relaxation delays between sequence repetitions. The method has been shown to be useful for time‐efficient whole‐body B1 mapping and correction of T1 maps obtained using a variable FA technique in the presence of nonuniform RF excitation. Magn Reson Med 57:192–200, 2007.Keywords
This publication has 28 references indexed in Scilit:
- Saturated double‐angle method for rapid B1+ mappingMagnetic Resonance in Medicine, 2006
- Rapid high‐resolution T1 mapping by variable flip angles: Accurate and precise measurements in the presence of radiofrequency field inhomogeneityMagnetic Resonance in Medicine, 2006
- In vivo method for correcting transmit/receive nonuniformities with phased array coilsMagnetic Resonance in Medicine, 2005
- Assessment and correction of B1‐induced errors in magnetization transfer ratio measurementsMagnetic Resonance in Medicine, 2004
- Effects of end‐ring/shield configuration on homogeneity and signal‐to‐noise ratio in a birdcage‐type coil loaded with a human headMagnetic Resonance in Medicine, 2003
- Radiofrequency map of an NMR coil by imagingMagnetic Resonance Imaging, 1993
- Experimental determination of three-dimensional RF magnetic field distribution of nmr coilsJournal of Magnetic Resonance (1969), 1991
- Parameter relations for the Shinnar-Le Roux selective excitation pulse design algorithm (NMR imaging)IEEE Transactions on Medical Imaging, 1991
- Magnetic field mappingMagnetic Resonance in Medicine, 1988
- Precision mapping of the B1 field using the rotating-frame experimentJournal of Magnetic Resonance (1969), 1987