In vivo multiple‐mouse MRI at 7 Tesla
- 7 October 2005
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 54 (5) , 1311-1316
- https://doi.org/10.1002/mrm.20683
Abstract
We developed a live high‐field multiple‐mouse magnetic resonance imaging method to increase the throughput of imaging studies involving large numbers of mice. Phantom experiments were performed in 7 shielded radiofrequency (RF) coils for concurrent imaging on a 7 Tesla MRI scanner outfitted with multiple transmit and receive channels to confirm uniform signal‐to‐noise ratio and minimal ghost artifacts across images from the different RF coils. Grid phantoms were used to measure image distortion in different positions in the coils. The brains of 7 live mice were imaged in 3D in the RF coil array, and a second array of 16 RF coils was used to 3D image the whole bodies of 16 fixed, contrast agent‐perfused mice. The images of the 7 live mouse brains at 156 μm isotropic resolution and the 16 whole fixed mice at 100 μm isotropic resolution were of high quality and free of artifacts. We have thus shown that multiple‐mouse MRI increases throughput for live and fixed mouse experiments by a factor equaling the number of RF coils in the scanner. Magn Reson Med, 2005.Keywords
This publication has 15 references indexed in Scilit:
- Parallel imaging for NMR microscopy at 14.1 TeslaMagnetic Resonance in Medicine, 2005
- Hexagonal zero mode TEM coil: A single‐channel coil design for imaging multiple small animalsMagnetic Resonance in Medicine, 2005
- In vivo measurements of T1 relaxation times in mouse brain associated with different modes of systemic administration of manganese chlorideJournal of Magnetic Resonance Imaging, 2005
- Multiple mouse biological loading and monitoring system for MRIMagnetic Resonance in Medicine, 2004
- Assessment of motion gating strategies for mouse magnetic resonance at high magnetic fieldsJournal of Magnetic Resonance Imaging, 2004
- Micro-MRI at 11.7 T of a Murine Brain Tumor Model Using Delayed Contrast EnhancementMolecular Imaging, 2003
- Super‐parallel MR microscopeMagnetic Resonance in Medicine, 2003
- High‐resolution MRI with cardiac and respiratory gating allows for accurate in vivo atherosclerotic plaque visualization in the murine aortic archMagnetic Resonance in Medicine, 2003
- In vivo multiple-mouse imaging at 1.5 TMagnetic Resonance in Medicine, 2003
- Tracing Odor-Induced Activation in the Olfactory Bulbs of Mice Using Manganese-Enhanced Magnetic Resonance ImagingNeuroImage, 2002