Image analysis methods for assessing levels of image plane nonuniformity and stochastic noise in a magnetic resonance image of a homogeneous phantom
- 8 August 2000
- journal article
- Published by Wiley in Medical Physics
- Vol. 27 (8) , 1980-1994
- https://doi.org/10.1118/1.1286593
Abstract
Magnetic response image plane nonuniformity and stochastic noise are properties that greatly influence the outcome of quantitative magnetic resonance imaging (MRI) evaluations such as gel dosimetry measurements using MRI. To study these properties, robust and accurate image analysis methods are required. New nonuniformity level assessment methods were designed, since previous methods were found to be insufficiently robust and accurate. The new and previously reported nonuniformity level assessment methods were analyzed with respect to, for example, insensitivity to stochastic noise; and previously reported stochastic noise level assessment methods with respect to insensitivity to nonuniformity. Using the same image data, different methods were found to assess significantly different levels of nonuniformity. Nonuniformity levels obtained using methods that count pixels in an intensity interval, and obtained using methods that use only intensity values, were found not to be comparable. The latter were found preferable, since they assess the quantity intrinsically sought. A new method which calculates a deviation image, with every pixel representing the deviation from a reference intensity, was least sensitive to stochastic noise. Furthermore, unlike any other analyzed method, it includes all intensity variations across the phantom area and allows for studies of nonuniformity shapes. This new method was designed for accurate studies of nonuniformities in gel dosimetry measurements, but could also be used with benefit in quality assurance and acceptance testing of MRI, scintillation camera, and computer tomography systems. The stochastic noise level was found to be greatly method dependent. Two methods were found to be insensitive to nonuniformity and also simple to use in practice. One method assesses the stochastic noise level as the average of the levels at five different positions within the phantom area, and the other assesses the stochastic noise in a region outside the phantom areaKeywords
This publication has 12 references indexed in Scilit:
- Automatic quality assessment protocol for MRI equipmentMedical Physics, 1999
- A comparison of two methods for measuring the signal to noise ratio on MR imagesPhysics in Medicine & Biology, 1999
- ResponseMagnetic Resonance in Medicine, 1996
- II. Performance assessment and quality control in MRI by Eurospin test objects and protocolsMagnetic Resonance Imaging, 1993
- III. Trial of modifications to Eurospin MRI test objectsMagnetic Resonance Imaging, 1993
- A new dosimeter based on ferrous sulphate solution and agarose gelInternational Journal of Radiation Applications and Instrumentation. Part A. Applied Radiation and Isotopes, 1991
- Quality assurance methods and phantoms for magnetic resonance imaging: Report of AAPM nuclear magnetic resonance Task Group No. 1Medical Physics, 1990
- IV. Protocols and test objects for the assessment of MRI equipmentMagnetic Resonance Imaging, 1988
- Measurement of signal intensities in the presence of noise in MR imagesMedical Physics, 1985
- A signal-to-noise calibration procedure for NMR imaging systemsMedical Physics, 1984