Measurement of transfer factor during constant exhalation.
Open Access
- 1 November 1994
- Vol. 49 (11) , 1121-1126
- https://doi.org/10.1136/thx.49.11.1121
Abstract
BACKGROUND--Transfer factor of the lung for carbon monoxide (TLCO) was measured by a new method based on analysis of the ratio of the concentrations of carbon monoxide to an inert gas (methane) relative to lung volume during a constant exhalation. Since this new technique is based solely upon exhalation, anomalies associated with inspiration and breath holding do not affect results. Additionally, because prolonged breath holding is not required, measurements can readily be made in dyspnoeic patients. METHODS--Exhalation TLCO (TLCO,ex) was compared with the standard (Jones and Meade) 10 second breath holding TLCO (TLCO,bh) in 100 consecutive patients. Patients did not practise the exhalation manoeuvre prior to testing. RESULTS--The comparative results were very close; mean difference (bias) +/- standard deviation (precision) was 0.05 (0.84) mmol/min/kPa. The relation was equally strong in patients with severe pulmonary disease; for patients with FEV1 < 1.51 the mean difference was 0.21 (0.80) mmol/min/kPa. CONCLUSIONS--Since the results were essentially identical between the techniques, it seems that comparable pathophysiological factors affect TLCO during breath holding and constant exhalation. Constant exhalation may therefore be a useful alternative to the breath holding technique for clinical measurement of TLCO.Keywords
This publication has 21 references indexed in Scilit:
- Measurement of cardiac output by automated single-breath technique, and comparison with thermodilution and Fick methods in patients with cardiac diseaseThe American Journal of Cardiology, 1993
- Adjustment of DlCO for varying COHb, and alveolar PO2 using a theoretical adjustment equationRespiration Physiology, 1990
- Measurement of Cardiac Output by Alveolar Gas ExchangePublished by American Geophysical Union (AGU) ,1987
- Rest and Exercise Cardiac Output and Diffusing Capacity Assessed By a Single Slow Exhalation of Methane, Acetylene, and Carbon MonoxideChest, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Non-invasive measurement of cardiac output by a single breath constant expiratory technique.Thorax, 1984
- Theoretical basis of single breath gas absorption testsJournal of Mathematical Biology, 1982
- Spirometric Studies in Normal SubjectsActa Medica Scandinavica, 1963
- PULMONARY DIFFUSING CAPACITY: AN IMPROVED SINGLE-BREATH METHODThe Lancet, 1960
- A Standardized Breath Holding Technique for the Clinical Measurement of the Diffusing Capacity of the Lung for Carbon Monoxide 1Journal of Clinical Investigation, 1957