An active microwave imaging system for reconstruction of 2-D electrical property distributions
- 1 January 1995
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 42 (10) , 1017-1026
- https://doi.org/10.1109/10.464376
Abstract
The goal of this work is to develop a microwave-based imaging system for hyperthermia treatment monitoring and assessment. Toward this end, a four transmit channel and four receive channel hardware device and concomitant image reconstruction algorithm have been realized. The hardware is designed to measure electric fields (i.e., amplitude and phase) at various locations in a phantom tank with and without the presence of various heterogeneities using standard heterodyning principles. Particular attention has been paid to designing a receiver with better than 115 dB of linear dynamic range which is necessary for imaging biological tissue which often has very high conductivity, especially for tissues with high water content. A calibration procedure has been developed to compensate for signal loss due to three-dimensional radiation in the measured data, since the reconstruction process is only two-dimensional at the present time. Results are shown which demonstrate the stability and accuracy of the measurement system, the extent to which the forward computational model agrees with the measured field distribution when the electrical properties are known, and image reconstructions of electrically unknown targets of varying diameter. In the latter case, images of both the reactive and resistive component of the electrical property distribution have been recoverable. Quantitative information on object location, size, and electrical properties results when the target is approximately one-half wavelength in size. Images of smaller objects lack the same level of quantitative information, but remain qualitatively correct.Keywords
This publication has 15 references indexed in Scilit:
- A dual mesh scheme for finite element based reconstruction algorithmsIEEE Transactions on Medical Imaging, 1995
- Temperature field estimation using electrical impedance profiling methods. I. Reconstruction algorithm and simulated resultsInternational Journal of Hyperthermia, 1994
- Weighted least-squares criteria for electrical impedance tomographyIEEE Transactions on Medical Imaging, 1992
- Memory and operations count scaling of coupled finite‐element and boundary‐element systems of equationsInternational Journal for Numerical Methods in Engineering, 1992
- Planar and cylindrical active microwave temperature imaging: numerical simulationsIEEE Transactions on Medical Imaging, 1992
- Medical imaging with a microwave tomographic scannerIEEE Transactions on Biomedical Engineering, 1990
- In Situ Permittivity of Canine Brain: Regional Variations and Postmortem ChangesIEEE Transactions on Microwave Theory and Techniques, 1986
- Limitations of Imaging with First-Order Diffraction TomographyIEEE Transactions on Microwave Theory and Techniques, 1984
- RF magnetic field penetration, phase shift and power dissipation in biological tissue: implications for NMR imagingPhysics in Medicine & Biology, 1978
- Nomographs for phase centers of conical corrugated and TE11mode hornsIEEE Transactions on Antennas and Propagation, 1975