Measurements of backscattered radiation from Therac‐20 collimator and trimmer jaws into beam monitor chamber
- 1 March 1989
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 16 (2) , 292-294
- https://doi.org/10.1118/1.596380
Abstract
The field size dependent photon output is known to be influenced by the existence of backscattered radiation (BSR) generated in the collimator or trimmer jaws of a linear accelerator. This paper describes the results of measurements made to study the existence of such backscatter by simulating the geometry of the treament head of a Therac-20 linear accelerator. The machine''s monitor chamber, flattening filter, and collimator jaws were simulated by another real monitor chamber, a 1-cm thick lead sheet and 2.5-cm thick low-melting-point alloy divergent blocks, respectively. BSR from the simulated collimator jaws (SCJ) was measured with the simulated monitor chamber (SMC) as a function of the openings of the SCJ and as a function of distance between SMC and SCJ. The present results demonstrate the presence of BSR in an 18-MV proton beam from a Therac-20 linear accelerator.This publication has 5 references indexed in Scilit:
- Telescopic measurements of backscattered radiation from secondary collimator jaws to a beam monitor chamber using a pair of slitsMedical Physics, 1989
- The effect of collimator backscatter radiation on photon output of linear acceleratorsMedical Physics, 1987
- Head scatter data for several linear accelerators (4-18 MV)Medical Physics, 1986
- Characteristics of an 18 MV photon beam from a Therac 20 Medical Linear AcceleratorMedical Physics, 1981
- Backscatter from High Atomic Number Materials in High Energy Photon BeamsRadiology, 1977