Telescopic measurements of backscattered radiation from secondary collimator jaws to a beam monitor chamber using a pair of slits
- 1 March 1989
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 16 (2) , 295-298
- https://doi.org/10.1118/1.596381
Abstract
A contribution to field-size dependent output by backscattered radiation (BSR) from secondary collimator jaws to a beam monitor chamber of a linear accelerator was measured with a Farmer ionization chamber, positioned 200 cm from the source behind a low-melting-point alloy slab with a 10-cm wide slit. Another slit was poitioned against the collimator jaws. Both slits were in the form of a 6.3-mm-diam hole in the middle and were aligned to the source. The use of a pair of slits was intended to eliminate any influence on the ion chamber readings due to field-size dependent charge contribution from the flattening filter and collimator jaw forward scattering. In addition, the setup permits to observe the degree of field-size dependence on BSR. Charge measurements from the Therac-20 18-MV x rays showed a 7.5% field-size dependence on BSR whereas 6- and 18-MV x rays from Varian Clinac-1800 showed <2% dependence on BSR. The telescopic method was found to be easy to use and permitted direct determination of BSR contributions.This publication has 6 references indexed in Scilit:
- Measurements of backscattered radiation from Therac‐20 collimator and trimmer jaws into beam monitor chamberMedical Physics, 1989
- Output factor constituents of a high‐energy photon beamMedical Physics, 1988
- Monte Carlo simulation of a cobalt‐60 beamMedical Physics, 1987
- The effect of collimator backscatter radiation on photon output of linear acceleratorsMedical Physics, 1987
- Head scatter data for several linear accelerators (4-18 MV)Medical Physics, 1986
- Characteristics of an 18 MV photon beam from a Therac 20 Medical Linear AcceleratorMedical Physics, 1981