Evaluation of an Automated Sample Preparation Protocol for Quantitative Detection of Hepatitis C Virus RNA
Open Access
- 1 April 2002
- journal article
- research article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 40 (4) , 1447-1450
- https://doi.org/10.1128/jcm.40.4.1447-1450.2002
Abstract
The COBAS AMPLIPREP instrument for automated sample preparation has recently been introduced. In this study, the COBAS AMPLIPREP/COBAS AMPLICOR HCV MONITOR test, which includes this new molecular device, was evaluated and compared to the COBAS AMPLICOR HCV MONITOR test, which includes a manual extraction protocol. Interassay and intra-assay variation, precision, and linearity were determined, and a total of 130 clinical specimens were investigated. For determination of interassay variation, coefficients of variation were found to be between 9 and 59% for the COBAS AMPLIPREP/COBAS AMPLICOR HCV MONITOR test and between 13 and 69% for the COBAS AMPLICOR HCV MONITOR test. For determination of intra-assay variation, coefficients of variation were found to be between 7 and 13% for the COBAS AMPLIPREP/COBAS AMPLICOR HCV MONITOR test and between 8 and 16% for the COBAS AMPLICOR HCV MONITOR test. When precision of the COBAS AMPLIPREP/COBAS AMPLICOR HCV MONITOR test was tested, all results were found to be within ±0.5 log of the expected results. Determination of linearity resulted in a quasilinear curve over 3 logs. When clinical samples were tested with the COBAS AMPLIPREP/COBAS AMPLICOR HCV MONITOR test and compared with the COBAS AMPLICOR HCV MONITOR test, all results were found within ±0.5 log. In conclusion, the assay, which included the new molecular device, proved to be suitable for the routine molecular laboratory. It was found to be laborsaving and easy to handle.Keywords
This publication has 18 references indexed in Scilit:
- Automation of laboratory testing for infectious diseases using the polymerase chain reaction — our past, our present, our futureJournal of Clinical Virology, 2001
- Evaluation of molecular parameters for routine assessment of viremia in patients with chronic hepatitis C who are undergoing antiviral therapy.1999
- Multicenter International Work Flow Study of an Automated Polymerase Chain Reaction InstrumentClinical Chemistry, 1998
- Evaluation of the automated COBAS AMPLICOR hepatitis C virus PCR systemJournal of Clinical Microbiology, 1997
- Performance of the automated COBAS AMPLICOR™ system for the detection of hepatitis C virus RNAClinical and Diagnostic Virology, 1997
- Comparison of QIAamp HCV kit spin columns, silica beads, and phenol-chloroform for recovering human immunodeficiency virus type 1 RNA from plasmaJournal of Clinical Microbiology, 1997
- Comparison of hepatitis B virus DNA extractions from serum by the QIAamp blood kit, GeneReleaser, and the phenol-chloroform methodJournal of Clinical Microbiology, 1996
- Evaluation of automated COBAS AMPLICOR PCR system for detection of several infectious agents and its impact on laboratory managementJournal of Clinical Microbiology, 1996
- The polymerase chain reaction, a review of the practical limitations for human immunodeficiency virus diagnosisJournal of Virological Methods, 1989
- Avoiding false positives with PCRNature, 1989