Comparison of flow artifacts between 2D-FLAIR and 3D-FLAIR sequences at 3 T
- 19 June 2004
- journal article
- research article
- Published by Springer Nature in European Radiology
- Vol. 14 (10) , 1901-1908
- https://doi.org/10.1007/s00330-004-2372-7
Abstract
It has been reported that 3D-FLAIR can reduce the flow artifact resulting from cerebrospinal fluid (CSF) at 1.5 T compared to 2D-FLAIR. Flow-related artifacts tend to be worse at 3 T than at 1.5 T. The purpose of this study was to compare the CSF flow artifacts of 2D-FLAIR and 3D-FLAIR sequences at 3 T in eight healthy volunteers. The grade of CSF-related artifacts were scored through observing the perimedullary cistern, cerebellopontine angle cisterns, fourth ventricule, prepontine cistern, suprasellar cistern, ambient cisterns, sylvian fissures, third ventricle and lateral ventricles. Grading was performed on either axial or sagittal images. The CSF in-flow artifact scores were significantly higher on axial 2D-FLAIR than on axial 3D-FLAIR MPR images in all areas except the bilateral sylvian fissures, and higher on sagittal 2D-FLAIR than on sagittal 3D-FLAIR MPR images in perimedullary, bilateral CP angle and suprasellar cisterns. The CSF-related flow artifacts were significantly reduced by 3D-FLAIR, while structures in the cistern were depicted more clearly, even at 3 T. Further study is necessary to compare the clinical efficacy between 2D-FLAIR and 3D-FLAIR in depicting subtle abnormalities.Keywords
This publication has 27 references indexed in Scilit:
- High field human imagingJournal of Magnetic Resonance Imaging, 2003
- Magnetic resonance imaging protocols for examination of the neurocranium at 3 TEuropean Radiology, 2003
- Abnormal hyperintensity within the subarachnoid space evaluated by fluid-attenuated inversion-recovery MR imaging: a spectrum of central nervous system diseasesEuropean Radiology, 2003
- Serial Isotropic Three-Dimensional Fast FLAIR Imaging: Using Image Registration and Subtraction to Reveal Active Multiple Sclerosis LesionsAmerican Journal of Roentgenology, 2002
- Generalized autocalibrating partially parallel acquisitions (GRAPPA)Magnetic Resonance in Medicine, 2002
- FLAIR imaging using nonselective inversion pulses combined with slice excitation order cycling andk-space reordering to reduce flow artifactsMagnetic Resonance in Medicine, 2001
- Optimized Single-Slab Three-dimensional Spin-Echo MR Imaging of the BrainRadiology, 2000
- Fast fluid-attenuated inversion-recovery imaging: first experience with a 3D version in epilepsyNeuroradiology, 1998
- Low-field versus high-field MR imaging of the knee: a comparison of signal behaviour and diagnostic performanceEuropean Journal of Radiology, 1995
- Use of Fluid Attenuated Inversion Recovery (FLAIR) Pulse Sequences in MRI of the BrainJournal of Computer Assisted Tomography, 1992