Modification of Rapid Human Immunodeficiency Virus (HIV) Antibody Assay Protocols for Detecting Recent HIV Seroconversion

Abstract
Assay protocols of three rapid human immunodeficiency virus (HIV) assays, OraQuick-1/2, SeroStrip-1/2, and Determine-1/2, were modified to detect recent HIV seroconversion using a higher dilution of serum specimens. Optimal predilution of specimens resulted in negative test results during early periods of seroconversion (about 6 months), when antibody levels were low. A total of 269 seropositive specimens from routine HIV type 1 testing and from commercial sources (low-titer and seroconversion panels) were tested, and results were recorded as negative (score = 0) or positive using intensity scores from 0.5 (weak positive) to 4 (strongly positive). The same specimens were previously tested by a less sensitive (LS) enzyme immunoassay (EIA), Abbott 3A11-LS, and were classified as recent or long-term infections based on the standardized optical density (SOD) cutoff of 0.75. Overall concordance of >94% was observed between 3A11-LS and modified rapid tests (RT-LSs) for detecting and distinguishing recent HIV seroconversion from long-term HIV infection (kappa statistics = 0.894 to 0.901). Moreover, intensity scores on RT-LSs correlated well with median 3A11-LS SOD values ( R 2 > 0.98). Our results indicate that rapid HIV tests can be modified to detect recent seroconversion with results comparable to those from less sensitive EIA.

This publication has 21 references indexed in Scilit: