Continuous monitoring of jugular bulb oxygen saturation as a measure of the shunt flow of cerebral arteriovenous malformations

Abstract
✓ Jugular bulb oxygen saturation (SjO2) was monitored during preoperative embolization procedures in a consecutive series of 15 patients with large supratentorial arteriovenous malformations (AVM's) in order to test the hypothesis that changes in the shunt flow ratio can be continuously evaluated from the SjO2. A fiberoptic catheter was placed at the dominant jugular bulb. The SjO2 measured using jugular blood withdrawn before embolization was significantly higher than the SjO2 measured at the end of the final embolization procedure (mean ± standard deviation 84.1% ± 12.7% vs. 74.2% ± 10.9%, p < 0.0001), showing a positive correlation with the AVM volume (r = 0.66, p < 0.001). Continuous monitoring of SjO2 via the fiberoptic catheter revealed a progressive decrease in association with the embolization procedures. Microsurgical resection of the AVM was performed at 1 to 2 weeks after the final embolization. Cases in which postoperative hemispheric deformation was revealed on computerized tomography demonst...