Investigation of C‐Arm Cone‐Beam CT‐Guided Surgery of the Frontal Recess
- 1 December 2005
- journal article
- research article
- Published by Wiley in The Laryngoscope
- Vol. 115 (12) , 2138-2143
- https://doi.org/10.1097/01.mlg.0000180759.52082.45
Abstract
Objective/Hypothesis: A cone‐beam CT (CBCT) imaging system based on a mobile C‐arm (Siemens PowerMobil) incorporating a high‐performance flat‐panel detector (Varian PaxScan) has been developed in our laboratory. We hypothesize that intraoperative C‐arm CBCT provides image quality and guidance performance sufficient to assist surgical approach to the frontal recess.Study Design: A preclinical prospective study was conducted using six cadaver heads to assess the performance characteristics and the potential clinical utility of this imaging system.Methods: The mobile C‐arm was employed for intraoperative CBCT guidance of the endoscopic approach to twelve frontal recesses.Results: The imaging system is capable of sub‐mm 3D spatial resolution with bone and soft‐tissue visibility and a field of view sufficient for guidance of head and neck surgery. The system can generate intraoperative, volumetric CT images rapidly with an acceptably low radiation exposure to the patient and with image quality sufficient for most surgical tasks. Moreover, the system is portable and compatible with the surgical setup, providing excellent access to the patient. Finally, the accuracy of the system is not bound to a registration process.Conclusions: The ability to create updated images as surgery progresses introduces the concept of ‘near‐real‐time’ CT guidance for head and neck surgery. We found that the use of CBCT increased surgical confidence in accessing the frontal recess, resolved ambiguities with anatomical variations, and provided valuable teaching information to surgeons in training in both preoperative planning and correlation between tri‐planar CT scans and intraoperative endoscopic findings.Keywords
This publication has 13 references indexed in Scilit:
- Volume CT with a flat‐panel detector on a mobile, isocentric C‐arm: Pre‐clinical investigation in guidance of minimally invasive surgeryMedical Physics, 2005
- Sources of Registration Error with Image Guidance Systems During Endoscopic Anterior Cranial Base SurgeryPublished by Wiley ,2004
- 3D Computed Tomographic Analysis of Frontal Recess Anatomy in Patients Without Frontal SinusitisOtolaryngology -- Head and Neck Surgery, 2004
- Impact of Image Guidance on Complications During Osteoplastic Frontal Sinus SurgeryOtolaryngology -- Head and Neck Surgery, 2004
- Image-guided surgery of the sinuses: current technology and applicationsOtolaryngologic Clinics of North America, 2004
- Revision Endoscopic Frontal Sinus Surgery with Surgical NavigationOtolaryngology -- Head and Neck Surgery, 2004
- The agger nasi cell: the key to understanding the anatomy of the frontal recessOtolaryngology -- Head and Neck Surgery, 2003
- Three-dimensional CT with a Modified C-Arm Image Intensifier: FeasibilityRadiology, 2002
- Multiplanar Reconstructed Computed Tomography Images Improves Depiction and Understanding of the Anatomy of the Frontal Sinus and RecessAmerican Journal of Rhinology, 2002
- Practical cone-beam algorithmJournal of the Optical Society of America A, 1984