Real‐time autoshimming for echo planar timecourse imaging
Open Access
- 1 November 2002
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 48 (5) , 771-780
- https://doi.org/10.1002/mrm.10259
Abstract
Head motion within an applied magnetic field alters the effective shim within the brain, causing geometric distortions in echo planar imaging (EPI). Even if subtle, change in shim can lead to artifactual signal changes in timecourse EPI acquisitions, which are typically performed for functional MRI (fMRI) or diffusion tensor imaging. Magnetic field maps acquired before and after head motions of clinically realistic magnitude indicate that motion‐induced changes in magnetic field may cause translations exceeding 3 mm in the phase‐encoding direction of the EPI images. The field maps also demonstrate a trend toward linear variations in shim changes as a function of position within the head, suggesting that a real‐time, first‐order correction may compensate for motion‐induced changes in magnetic field. This article presents a navigator pulse sequence and processing method, termed a “shim NAV,” for real‐time detection of linear shim changes, and a shim‐compensated EPI pulse sequence for dynamic correction of linear shim changes. In vivo and phantom experiments demonstrate the detection accuracy of shim NAVs in the presence of applied gradient shims. Phantom experiments demonstrate reduction of geometric distortion and image artifact using shim‐compensated EPI in the presence of applied gradient shims. In vivo experiments with intentional interimage subject motion demonstrate improved alignment of timecourse EPI images when using the shim NAV‐detected values to update the shim‐compensated EPI acquisition in real time. Magn Reson Med 48:771–780, 2002.Keywords
This publication has 20 references indexed in Scilit:
- Real-time auto-shimming for FMRINeuroImage, 2001
- Compensation of susceptibility induced BOLD sensitivity losses in echo-planar fMRI imagingNeuroImage, 2001
- A prospective approach to correct for inter‐image head rotation in FMRIMagnetic Resonance in Medicine, 1998
- Inadequacy of motion correction algorithms in functional MRI: Role of susceptibility‐induced artifactsJournal of Magnetic Resonance Imaging, 1997
- AFNI: Software for Analysis and Visualization of Functional Magnetic Resonance NeuroimagesComputers and Biomedical Research, 1996
- Movement‐Related effects in fMRI time‐seriesMagnetic Resonance in Medicine, 1996
- Orbital navigator echoes for motion measurements in magnetic resonance imagingMagnetic Resonance in Medicine, 1995
- Artifacts due to stimulus correlated motion in functional imaging of the brainMagnetic Resonance in Medicine, 1994
- Analysis of T2 limitations and off‐resonance effects on spatial resolution and artifacts in echo‐planar imagingMagnetic Resonance in Medicine, 1990