Evaluation of the Partec Flow Cytometer against the BD FACSCalibur System for Monitoring Immune Responses of Human Immunodeficiency Virus-Infected Patients in Zimbabwe
- 1 March 2007
- journal article
- Published by American Society for Microbiology in Clinical and Vaccine Immunology
- Vol. 14 (3) , 293-298
- https://doi.org/10.1128/cvi.00416-06
Abstract
A single-platform volumetric flow cytometer, the Partec Cyflow SL_3, was evaluated against a BD FACSCalibur/Sysmex XT1800i dual platform for measuring CD4(+) lymphocytes, total lymphocytes, and the percentage of CD4 lymphocytes in whole-blood samples for monitoring the immune systems of human immunodeficiency virus (HIV)/AIDS patients. Statistical analyses for precision, correlation, and agreement were performed. Coefficients of variation (CV) of 5.8, 4.6, and 3.9% were obtained for low, medium, and high CD4(+) cell counts, respectively, using the SL_3, and CV of 3.7, 4.0, and 0.94 were obtained for the same categories, using the BD FACSCalibur. Significant correlations (P < 0.005) between the two assays for CD4 counts, total lymphocyte counts, and percentages of CD4 were obtained, with correlation coefficients of 0.99, 0.96, and 0.99, respectively (n = 229). Using the Bland-Altman plot, mean biases of -18 cell/microl (95% confidence interval (CI); -91 to 54 cells/microl), -0.8% (95% CI; -3.6 to 2%), and -36.8 cells/microl (95% CI; -477 to 404 cells/microl) were obtained for comparisons of CD4 counts, percentages of CD4 cells, and total lymphocyte counts, respectively. The effects of the age of the samples on the three parameters were also analyzed by comparing results from the same samples analyzed at 6, 24, and 48 h after collection. The correlation coefficients for comparisons among different time points for the same machine and among all the time points for the two different machines were greater than 0.90. These data showed that the Partec Cyflow SL_3 assay is comparable to the BD FACSCalibur/Sysmex XT1800i dual-platform method for measuring the amount of CD4(+) cells and total lymphocytes and the percentages of CD4 cells in blood samples for the purpose of monitoring HIV/AIDS patients.Keywords
This publication has 14 references indexed in Scilit:
- Comparison of two alternative methods for CD4+ T‐cell determination (Coulter manual CD4 count and CyFlow) against standard dual platform flow cytometry in UgandaCytometry Part B: Clinical Cytometry, 2006
- Evaluation of a New Single-Parameter Volumetric Flow Cytometer (CyFlow green ) for Enumeration of Absolute CD4 + T Lymphocytes in Human Immunodeficiency Virus Type 1-Infected Thai PatientsClinical and Vaccine Immunology, 2005
- Generic HIV Drugs — Enlightened Policy for Global HealthNew England Journal of Medicine, 2005
- Comparison of a New, Affordable Flow Cytometric Method and the Manual Magnetic Bead Technique for CD4 T-Lymphocyte Counting in a Northern Nigerian SettingClinical and Vaccine Immunology, 2005
- Simplified Volumetric Flow Cytometry Allows Feasible and Accurate Determination of Cd4 T Lymphocytes in Immunodeficient Patients WorldwideAntiviral Therapy, 2004
- A New No-Lyse, No-Wash Flow-Cytometric Method for the Determination of CD4 T Cells in Blood SamplesTransfusion Medicine and Hemotherapy, 2003
- Health-Related Quality of Life After 1 Year of Highly Active Antiretroviral TherapyJAIDS Journal of Acquired Immune Deficiency Syndromes, 2003
- Evaluation of TruCount Absolute-Count Tubes for Determining CD4 and CD8 Cell Numbers in Human Immunodeficiency Virus-Positive AdultsClinical and Diagnostic Laboratory Immunology, 2000
- Rapid turnover of plasma virions and CD4 lymphocytes in HIV-1 infectionNature, 1995
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986