Ex vivo evaluation of ferromagnetism, heating, and artifacts produced by heart valve prostheses exposed to a 1.5‐T MR system
- 1 September 1994
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 4 (5) , 756-758
- https://doi.org/10.1002/jmri.1880040521
Abstract
Ex vivo testing techniques were used to determine the ferromagnetic qualities of, presence of heating in, and artifacts produced by 13 different heart valve prostheses exposed to a 1.5‐T (64‐MHz) magnetic resonance (MR) system. None of the heart valve prostheses showed a measurable deflection in the 1.5‐T static magnetic field. Only minimal artifacts were produced during MR imaging with a fast spoiled GRASS (gradient‐recalled acquisition in the steady state) pulse sequence. The largest temperature changes measured during a “worst case” MR imaging sequence (estimated average specific absorption rate, 2.5 W/kg; estimated spatial peak specific absorption rate, 7.6 W/kg) were +0.2°C with the implant imaged “in air” and +0.3°C with the implant imaged in normal saline. Therefore, MR procedures performed with a 1.5‐T (64‐MHz) MR system may be performed safely in patients with any of the 13 different heart valve prostheses evaluated in this study.Keywords
This publication has 16 references indexed in Scilit:
- MR procedures and biomedical implants, materials, and devices: 1993 update.Radiology, 1993
- MR imaging and biomedical implants, materials, and devices: an updated review.Radiology, 1991
- Important considerations in measurement of attractive force on metallic implants in MR imagers.Radiology, 1991
- Safety considerations in MR imaging.Radiology, 1990
- MR imaging of metallic implants and materials: a compilation of the literatureAmerican Journal of Roentgenology, 1988
- High-field-strength MR imaging and metallic biomedical implants: an ex vivo evaluation of deflection forcesAmerican Journal of Roentgenology, 1988
- MR imaging artifacts, ferromagnetism, and magnetic torque of intravascular filters, stents, and coils.Radiology, 1988
- Magnetic resonance imaging artifacts: mechanism and clinical significance.RadioGraphics, 1986
- Magnetic resonance imaging of prosthetic heart valves.Radiology, 1985
- Potential hazards in NMR imaging: heating effects of changing magnetic fields and RF fields on small metallic implantsAmerican Journal of Roentgenology, 1981