Enzyme-Multiplied Immunoassay (EMIT 2000) Digoxin Assay Compared with Fluorescence Polarization Immunoassay and Amerlex 125I-Radioimmunoassay at Two Australian Centers
- 1 December 1996
- journal article
- research article
- Published by Wolters Kluwer Health in Therapeutic Drug Monitoring
- Vol. 18 (6) , 672-677
- https://doi.org/10.1097/00007691-199612000-00008
Abstract
Digoxin assays in plasma from patients treated with the drug have played an integral role in its therapeutic management. Commercial digoxin immunoassays have been criticized for poor performance owing to various interferences and limited sensitivity. The present study compared the performance of a new enzyme-multiplied immunoassay technique (EMIT 2000) to fluorescence polarization immunoassay (FPIA) and radioimmunoassay (RIA) in two separate Australian centers. Comparisons were made using standard indices of precision and accuracy, samples taken from patients, quality-assurance samples, and cord blood samples from neonates, in which high concentrations of digoxin-like immunoreactive substances (DLIS) would be anticipated. The results confirmed satisfactory precision and accuracy for therapeutic drug monitoring purposes, a sensitivity of < 0.1 microgram/L, and very low DLIS interference, as assessed by assay of neonatal cord blood samples.Keywords
This publication has 10 references indexed in Scilit:
- A New Digoxin Immunoassay Substantially Free of Interference by Digoxin Immunoreactive FactorTherapeutic Drug Monitoring, 1995
- Treatment of Human Serum with Sulfosalicylic Acid Structurally Alters Digoxin and Endogenous Digoxin-Like Immunoreactive FactorTherapeutic Drug Monitoring, 1995
- Evaluation and Comparison of the TDXII, Stratus, and OPUS Digoxin AssaysTherapeutic Drug Monitoring, 1995
- Highly Specific Radioimmunoassay for Digoxin Using a Monoclonal Antibody Selected for Lack of Interference by Digoxin-Like Immunoreactive Substances in Cord Blood SeraTherapeutic Drug Monitoring, 1991
- Reassessment of Cross-Reactivity of Spironolactone Metabolites with Four Digoxin ImmunoassaysTherapeutic Drug Monitoring, 1989
- Digoxin-like substance in the serum of uremic patients before and after hemodialysisCardiovascular Drugs and Therapy, 1988
- Falsely raised plasma digoxin concentrations in liver disease.BMJ, 1985
- Falsely Increased Digoxin Concentrations in Samples from Neonates and InfantsTherapeutic Drug Monitoring, 1984
- Endogenous substance in newborn infants causingfalse positive digoxin measurementsThe Journal of Pediatrics, 1983
- Plasma digoxin concentration--its relation to digoxin dosage and clinical effects in patients with atrial fibrillation.Heart, 1972