Comparison of portable automated keratometry and manual keratometry for IOL calculation
- 1 October 1997
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 23 (8) , 1213-1216
- https://doi.org/10.1016/s0886-3350(97)80318-5
Abstract
Purpose: To compare the accuracy of portable automated keratometry (PAK) with that of manual keratometry (MK) in measuring corneal power for intraocular lens calculations. Setting: Wilford Hall Medical Center, San Antonio, Texas, USA. Methods: In Part 1 of the study, five ophthalmic technicians performed keratometric analysis in 20 eyes in 10 volunteers using both manual and automated methods to determine the relative accuracy and reproducibility of each instrument. In Part 2, both MK and PAK were prospectively performed in 11 patients having cataract surgery to compare the accuracy of each instrument in predicting postoperative refractive outcome. Results: The difference between instruments in determining the average corneal power in all eyes was less than 0.10 diopter (D) (MK = 43.84 D, PAK=43.93 D). Portable keratometry demonstrated less variability in measurements in each eye (average standard deviation, MK=0.30 D, PAK=0.11 D; average range, MK=1.08 D, PAK=0.44 D). The mean absolute refractive error (difference between the actual refractive outcome and predicted refractive outcome) was 0.37 D ± 0.30 (SD) using MK values and 0.45 ± 0.19 D using PAK values. Conclusions: Portable automated keratometry is a simple keratometric technique that appeared to be as accurate as but with less variability than manual keratometry in determining corneal power for cataract surgery.Keywords
This publication has 6 references indexed in Scilit:
- Comparison of three keratometry instrumentsJournal of Cataract & Refractive Surgery, 1995
- Comparison of the accuracy of computerized videokeratography and keratometry for use in intraocular lens calculationsJournal of Cataract & Refractive Surgery, 1993
- Comparison of the accuracy and reproducibility of the keratometer and the EyeSys Corneal Analysis System Model IJournal of Cataract & Refractive Surgery, 1992
- Sources of error in intraocular lens power calculationJournal of Cataract & Refractive Surgery, 1992
- Comparison of the SRK II™ formula and other second generation formulasJournal of Cataract & Refractive Surgery, 1988
- Factors Associated With Poor Predictability by Intraocular Lens Calculation FormulasArchives of Ophthalmology (1950), 1985