Implant provision of key, pinch and power grips in a C6 tetraplegic
- 1 July 1994
- journal article
- case report
- Published by Springer Nature in Medical & Biological Engineering & Computing
- Vol. 32 (4) , 367-372
- https://doi.org/10.1007/bf02524686
Abstract
An 11-channel multiplexed stimulator of nerves and muscles in the left forearm was implanted for hand control in January 1986 in a 21 year old woman who, after sustaining a C6 spinal lesion 7 years earlier, had voluntary shoulder and elbow movement but paralysed hands, trunk and legs. The patient controls the stimulation via a microcomputer control box and an RF transdermal link. We have investigated the control of her stimulated hand with a joystick under her contralateral hand which she moves from the shoulder and elbow. Since 1986, we have tried a variety of joystick control schemes involving power and key grips. Currently, for grip adjustment, forward and backward joystick movements correspond to thumb extension and abduction, respectively giving in addition both finger and wrist extension, whereas right and left joystick movements yield first closure and thumb opposition/adduction and flexion, respectively. Useful grasps are available by moving the joystick forward and then left (key grip), by moving the joystick backward and left (pinch grip), or by moving the joystick back and right (power grip). Thus, three distinct grips may be selected using these three quadrants of joystick movement. An additional control mode was found to be desirable to augment the patient's limited voluntary wrist positioning and provide wrist stability while adjusting finger grip.Keywords
This publication has 7 references indexed in Scilit:
- FNS of the upper limb: targeting the forearm muscles for surface stimulationMedical & Biological Engineering & Computing, 1990
- Evaluation of shoulder movement as a command control sourceIEEE Transactions on Biomedical Engineering, 1990
- Functional neuromuscular stimulation neuroprostheses for the tetraplegic hand.1988
- Functional Neuromuscular Stimulation Neuroprostheses for the Tetraplegic HandPublished by Wolters Kluwer Health ,1988
- When are actively balanced biphasic (‘Lilly’) stimulating pulses necessary in a neurological prosthesis? II pH changes; noxious products; electrode corrosion; discussionMedical & Biological Engineering & Computing, 1986
- The Cooper cable: an implantable multiconductor cable for neurological prosthesesMedical & Biological Engineering & Computing, 1983
- Controlled prehension and release in the C5 quadriplegic elicited by functional electrical stimulation of the paralyzed forearm musculatureAnnals of Biomedical Engineering, 1980