Physical Characteristics and Clinical Evaluation of a New Disposable Fibreoptic Transducer-Tipped Catheter System
Open Access
- 1 August 1987
- journal article
- research article
- Published by SAGE Publications in Anaesthesia and Intensive Care
- Vol. 15 (3) , 323-329
- https://doi.org/10.1177/0310057x8701500313
Abstract
A new disposable fibreoptic transducer-tipped catheter manometer system was evaluated to assess its accuracy, stability of accuracy under prolonged simulated intra-arterial conditions, and dynamic characteristics. Maximum errors observed in the measurement of static pressure using a sample of five catheters (with one display unit) were 2 mmHg at 0 mmHg reference pressure, 2 at 20 mmHg, 4 at 40 mmHg, 4 at 100 mmHg and 9 at 200 mmHg. An immersion artifact caused a shift in baseline of up to 2 mmHg. Exposure of the transducer to 24 hours of simulated intra-arterial conditions (pulsatile pressure at 40°C) resulted in errors of up to 7 mmHg for pressures up to 100 mmHg, and 11 mmHg for 200 mmHg, which were largely attributable to a drift in baseline pressure (up 6 mmHg by 24 hours). Consistent overestimation by the system suggested inappropriate gain setting within the display unit which, however, is not user-adjustable. The system exhibited uniform frequency response up to 33 Hz.Keywords
This publication has 13 references indexed in Scilit:
- Continuous ambulatory pulmonary artery pressure monitoring. A new method using a transducer tipped catheter and a simple recording system.Heart, 1986
- Effect of the dynamic response of transducer-tubing system on accuracy of direct blood pressure measurement in patientsCritical Care Medicine, 1983
- An Evaluation of Blood Pressure MeasurementAnaesthesia and Intensive Care, 1981
- Direct Blood Pressure Measurement —Dynamic Response RequirementsAnesthesiology, 1981
- Artifactual Hypotension Without Damping, a Hazard of Disposable Diaphragm DomesAnesthesiology, 1979
- Laboratory Evaluation of Pressure Transducer Domes Containing a DiaphragmAnesthesia & Analgesia, 1978
- Analysis and Correction of Pressure Wave Distortion in Fluid-Filled Catheter SystemsCirculation, 1974
- Physical criteria for measurement of left ventricular pressure and its first derivativeCardiovascular Research, 1971
- Damped and undamped frequency responses of underdamped catheter manometer systemsAmerican Heart Journal, 1970
- Physiologic Recording by Modern Instruments With Particular Reference to Pressure RecordingPhysiological Reviews, 1960