MLE reconstruction of a brain phantom using a Monte Carlo transition matrix and a statistical stopping rule
- 1 February 1988
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 35 (1) , 603-607
- https://doi.org/10.1109/23.12795
Abstract
In order to study properties of the maximum-likelihood-estimator (MLE) algorithm for image reconstruction in positron-emission tomography (PET), the algorithm is applied to data obtained by the ECAT-III tomograph from a brain phantom. The procedure for subtracting accidental coincidences from the data stream generated by this physical phantom is such that the resultant data are not Poisson distributed. This makes the present investigation different from other investigations based on computer-simulated phantoms. It is shown that the MLE algorithm is robust enough to yield comparatively good images, especially when the phantom is in the periphery of the field of view, even though the underlying assumption of the algorithm is violated. A stopping rule derived earlier and allowing the user to stop the iterative process before the images begin to deteriorate is tested. Since the rule is based on the Poisson assumption, it does not work well with the presently available data, although it is successful with computer-simulated Poisson data.Keywords
This publication has 11 references indexed in Scilit:
- Possible Criteria for Choosing the Number of Iterations in Some Iterative Reconstruction MethodsPublished by Springer Nature ,1988
- Stopping Rule for the MLE Algorithm Based on Statistical Hypothesis TestingIEEE Transactions on Medical Imaging, 1987
- Noise and Edge Artifacts in Maximum-Likelihood Reconstructions for Emission TomographyIEEE Transactions on Medical Imaging, 1987
- The High Sensitivity of the Maximum Likelihood Estimator Method of Tomographic Image ReconstructionPublished by Springer Nature ,1987
- Inverse Monte Carlo: A Unified Reconstruction Algorithm for SPECTIEEE Transactions on Nuclear Science, 1985
- Design Analysis and Performance Evaluation of a Two Dimensional Camera for Accelerated Positron Emitter Beam Injection by Computer SimulationIEEE Transactions on Nuclear Science, 1983
- Maximum Likelihood Reconstruction for Emission TomographyIEEE Transactions on Medical Imaging, 1982
- An Imaging Instrument for Positron Emitting Heavy Ion Beam InjectionIEEE Transactions on Nuclear Science, 1979
- Design and Performance Characteristics of a Positron Emission Computed Axial Tomograph--ECAT®-IIIEEE Transactions on Nuclear Science, 1979
- Study of positronium in water and ice from 22 to -144 °C by annihilation quanta measurementsIl Nuovo Cimento (1869-1876), 1965