Polyethylene wear vector in vivo: A three‐dimensional analysis using retrieved acetabular components and radiographs
- 1 September 1999
- journal article
- research article
- Published by Wiley in Journal of Orthopaedic Research
- Vol. 17 (5) , 695-702
- https://doi.org/10.1002/jor.1100170511
Abstract
Polyethylene wear of the acetabular component can be described as one or more vectors. To help clarify the mechanisms of wear advancement in vivo, we used a combination of retrieved implants and radiographs to describe the three-dimensional wear vectors in total hip arthroplasty. The wear vectors in 41 retrieved implants from a single manufacturer were measured with use of the shadowgraph technique, and the spatial orientation of each implant was calculated from serial anteroposterior pelvic radiographs. On the basis of the combination of the wear vector in the implant and implant orientation in the pelvis, the wear vectors in vivo were determined. The mean wear vector was directed 8.1° lateral in the coronal plane and 4.1° posterior in the sagittal plane. The wear vector in vivo showed a relatively wide range of directions, not necessarily coinciding with the commonly presumed resultant force in the hip. The wear vectors were not associated with the spatial orientation of the acetabular components, but cups with impingement demonstrated more anterior wear than did those without impingement. Our results suggest that the process of polyethylene wear is not as simple as previously described and that several factors influence advancement of wear in vivo.Keywords
This publication has 30 references indexed in Scilit:
- Loci of movement of selected points on the femoral head during normal gaitThe Journal of Arthroplasty, 1996
- Effect of superior and superolateral relocations of the hip center on hip joint forcesThe Journal of Arthroplasty, 1996
- Adaptiive finite element modeling of long‐term polyethylene wear in total hip arthroplastyJournal of Orthopaedic Research, 1996
- Radiographic calculation of anteversion in acetabular prosthesesThe Journal of Arthroplasty, 1995
- Loosening of a femoral stem associated with the use of an extended-lip acetabular cup liner: A case reportThe Journal of Arthroplasty, 1993
- Polyethylene debris-induced osteolysis and loosening in uncemented total hip arthroplastyThe Journal of Arthroplasty, 1992
- The Mechanism of Loosening of Cemented Acetabular Components in Total Hip ArthroplastyClinical Orthopaedics and Related Research, 1992
- Wear of Hip SocketsPublished by Springer Nature ,1979
- Rate of Wear in Total Hip ReplacementClinical Orthopaedics and Related Research, 1975