Motion artifact reduction with fast spin-echo imaging.
- 1 July 1987
- journal article
- research article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 164 (1) , 183-191
- https://doi.org/10.1148/radiology.164.1.3588902
Abstract
The influence of signal averaging (n), repetition time (TR), and echo delay (TE) on systematic noise (cardiac, vascular, respiratory and peristaltic ghost artifacts) and statistical noise (thermal effects) was determined in eight healthy volunteers and 57 patients. Systematic noise was the dominant factor degrading abdominal magnetic resonance (MR) images. Signal averaging was the primary determinant of both statistical and systematic image noise, fitting a power function (n)b with b = -0.44 and -0.42, respectively, close to the expected b = -0.5 power function. All types of ghosting showed the same sensitivity to signal averaging. Normalized systematic noise increased slightly with TR (b = 0.16) and increased markedly with TE (b = 0.04). These data indicate that the short TR, short TE technique is a powerful method for reducing motion artifacts on breathhold images and can be combined with signal averaging to further suppress artifacts, improve signal-to-noise ratio, and maximize anatomic resolution.This publication has 17 references indexed in Scilit:
- Rapid NMR imaging of dynamic processes using the FLASII techniqueMagnetic Resonance in Medicine, 1986
- Liver metastases: detection by phase-contrast MR imaging.Radiology, 1986
- Magnetic resonance imaging of cavernous hemangioma of the liver: tissue-specific characterizationAmerican Journal of Roentgenology, 1985
- MR ImagingJournal of Computer Assisted Tomography, 1985
- Visualization and Measurement of Flow with Magnetic Resonance Imaging. Darstellung und Messung von Flußgeschwindigkeit mit bildgebender KernresonanzBiomedizinische Technik/Biomedical Engineering, 1985
- Magnetic resonance imaging with respiratory gating: techniques and advantagesAmerican Journal of Roentgenology, 1984
- Magnetic resonance imaging systems: optimization in clinical use.Radiology, 1984
- The effect of motion on two-dimensional Fourier transformation magnetic resonance images.Radiology, 1984
- Magnetic resonance imaging: effects of magnetic field strength.Radiology, 1984
- Cardiac imaging using gated magnetic resonance.Radiology, 1984