Construction and Calibration of a Cylindrical Electrostatic Analyzer
- 1 October 1949
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 20 (10) , 735-744
- https://doi.org/10.1063/1.1741368
Abstract
A cylindrical electrostatic deflector has been constructed with average radius 15 cm, spacing 0.5 cm and deflection 90°, in which the deflecting potential can be in the neighborhood of 50 kv. Thus protons up to 0.75 Mev can be focused. Movable slits have been incorporated in the design, and the geometry of the entrance and exit apertures of the deflecting channel is such that the stray field may be calculated analytically. This has been done for the case in which the deflecting plates are symmetrically charged above and below zero potential. The predictions have been compared with experiments on the focusing of electron beams by the instrument under controlled conditions. Agreement with the calculations is demonstrated within the accuracy of the experiments, which is about 0.3 percent. Experiments with asymmetric charging of the electrodes are reported and shown to agree with a plausible extension of the theory. Further experiments investigated the effect of small displacements of the slits from the computed focal positions. A set of factors, or calibration constants, has been measured so that, given the coordinates of the centers of the entrance and exit slits, and the potentials applied to the deflecting plates, the kinetic energy of the focused ion beam, as it left its source at zero potential, can be calculated to within 0.3 percent.Keywords
This publication has 19 references indexed in Scilit:
- A Direct Determination of the Energy of theNucleus from the D - D ReactionPhysical Review B, 1948
- Use of an Electron Multiplier Tube as a New Technique in Disintegration ExperimentsPhysical Review B, 1948
- An Independent Determination of Fixed Points on the High Voltage ScalePhysical Review B, 1944
- Corrections to Electrostatic Analyzer MeasurementsPhysical Review B, 1940
- A Precise Determination of the Energy Released in the Production of Deuterium from Beryllium under Proton BombardmentPhysical Review B, 1939
- A Precise Measurement of the Mass Difference—; The Stability ofPhysical Review B, 1939
- A Determination offrom the Short Wave-Length Limit of the Continuous X-Ray SpectrumPhysical Review B, 1937
- Electric and Magnetic Focusing in Mass SpectroscopyPhysical Review B, 1937
- Ablenkung von Kathoden- und Kanalstrahlen am Rande eines Kondensators, dessen Streufeld durch eine Blende begrenzt istThe European Physical Journal A, 1935
- Ionen- und elektronenoptische Zylinderlinsen und Prismen. IThe European Physical Journal A, 1934