Measurement of respiratory rate and timing using a nasal thermocouple
- 1 May 1995
- journal article
- research article
- Published by Springer Nature in Journal of Clinical Monitoring and Computing
- Vol. 11 (3) , 159-164
- https://doi.org/10.1007/bf01617716
Abstract
Objective. The aims of this study were to assess aspects of the response of a small thermocouple to temperature change, and to evaluate whether such a thermocouple could be used intermittently to measure respiratory rate and timing by detecting the changes in nasal temperature occurring with breathing.Methods. The study had three parts. First, three similar, fast-responding thermocouples were immersed repeatedly in warm water. Second, the influence of atmospheric temperature on the signal of a thermocouple placed at different sites within the nasal orifice was studied. The signals produced were continuously displayed and analyzed using a laptop computer to allow evaluation of the thermocouples' response characteristics. Third, simultaneous respiratory recordings were acquired using a nasal thermocouple and a nasal pneumotachograph in 12 teenaged subjects. The respiratory rate and the periods of time taken for inspiration (Ti) and expiration (Te) were calculated and compared.Results. The thermocouples' responses to the temperature changes associated with breathing and immersion into water were rapid and consistent. The rate of the signals' decay, following the peak signal marking expiration, was influenced by the atmospheric temperature. The time constants of the thermocouples were similar (mean time constant=0.41 sec, standard deviation (SD)=0.07). Optimal respiratory recordings were obtained, with least discomfort, when the thermocouple was positioned at 0 to 4 mm within the nasal orifice. In comparing the respiratory recordings acquired simultaneously with a thermocouple and pneumotachograph, the respiratory rates were identical, and the Ti and Te values were similar (mean difference 0.04 sec (95% CI: -0.11 to 0.21 sec) and -0.04 sec (95% CI: -0.20 to 0.12 sec), respectively).Conclusions. Intermittent measurements of respiratory rate and timing using a nasal thermocouple accurately reflected measurements obtained from nasal airflow using a pneumotachograph.Keywords
This publication has 14 references indexed in Scilit:
- Problems in Timing of Respiration With the Nasal Thermistor TechniqueJournal of the American Society of Echocardiography, 1993
- The Extensometer?use in spontaneously ventilating awake volunteersAnaesthesia, 1992
- Respiratory timing in intubated neonates with respiratory distress syndrome.Archives of Disease in Childhood, 1992
- Respiratory rate: measurement of variability over time and accuracy at different counting periods.Archives of Disease in Childhood, 1991
- Continuous Noninvasive Monitoring of Respiratory Rate in Critically III PatientsChest, 1986
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- BREATHING PATTERNS IN PATIENTS WITH CHRONIC OBSTRUCTIVE PULMONARY-DISEASEPublished by Elsevier ,1984
- Evaluation of two electronic respiratory rate monitoring systems.1981
- Ventilatory Pattern during Handgrip in Normal ManRespiration, 1981
- Some New Instruments for the Continuous Monitoring of Body Temperature, Respiration-Rate and Pulse-RatePhysics in Medicine & Biology, 1966