Catheter Intramural Needle Radiofrequency Ablation Creates Deeper Lesions than Irrigated Tip Catheter Ablation
- 30 October 2003
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 26 (11) , 2146-2150
- https://doi.org/10.1046/j.1460-9592.2003.00334.x
Abstract
Radiofrequency ablation of the left ventricle using an endocardially placed electrode is unable to reliably create transmural lesions even with active electrode cooling. To produce deeper radiofrequency lesions, the authors developed and tested a prototype intramural needle ablation catheter that had a distal 1.1‐mm diameter straight needle that could be advanced 12 mm into the myocardium. Freshly excised hearts from eight male sheep were perfused and superfused with oxygenated ovine blood. Ablations were performed for 60 seconds with the prototype catheter and a conventional 5‐mm irrigated tip ablation catheter at target temperatures of 90°C and 50°C, respectively. The ablation lesions were bisected and stained with blue tetrazolium to assess lesion geometry. The irrigated tip ablation catheter required significantly more power than the intramural needle ablation catheter (37.7 ± 7.3 vs 6.4 ± 2.1 W, P < 0.01). Intramural needle lesions were significantly deeper (12.5 ± 3.0 mm vs 8.3 ± 2.1 mm, P < 0.01) but less wide (3.9 ± 1.1 mm vs 11.5 ± 2.0 mm, P < 0.01) than irrigated tip lesions. There was a high incidence of crater formation (74%), popping (45%), and myocardial charring (29%) during irrigated tip ablation; these phenomena were not observed during intramural needle ablation. The intramural needle ablation catheter creates significantly deeper but narrower lesions without evidence of tissue boiling. This technology may be particularly useful for ablation of ventricular tachycardia originating from regions where tissue depth is increased, like the ventricular septum. (PACE 2003; 26:2146–2150)Keywords
This publication has 13 references indexed in Scilit:
- Linear Ablation Lesions for Control of Unmappable Ventricular Tachycardia in Patients With Ischemic and Nonischemic CardiomyopathyCirculation, 2000
- Prospective Randomized Comparison of Irrigated-Tip Versus Conventional-Tip Catheters for Ablation of Common FlutterCirculation, 2000
- A new catheter design using needle electrode for subendocardial RF ablation of ventricular muscles: finite element analysis and in vitro experimentsIEEE Transactions on Biomedical Engineering, 2000
- Lesion Dimensions During Temperature-Controlled Radiofrequency Catheter Ablation of Left Ventricular Porcine MyocardiumCirculation, 1999
- A cooled needle electrode for radiofrequency tissue ablation: Thermodynamic aspects of improved performance compared with conventional needle designAcademic Radiology, 1996
- Radiofrequency Delivery Through a Cooled Catheter Tip Allows the Creation of Larger Endomyocardial Lesions in the Ovine HeartJournal of Cardiovascular Electrophysiology, 1995
- Comparison of In Vivo Tissue Temperature Profile and Lesion Geometry for Radiofrequency Ablation With a Saline-Irrigated Electrode Versus Temperature Control in a Canine Thigh Muscle PreparationCirculation, 1995
- Development of a new intraoperative radiofrequency ablation technique using a needle electrodeThe Annals of Thoracic Surgery, 1994
- Functional role of the epicardium in postinfarction ventricular tachycardia. Observations derived from computerized epicardial activation mapping, entrainment, and epicardial laser photoablation.Circulation, 1991
- Electrode radius predicts lesion radius during radiofrequency energy heating. Validation of a proposed thermodynamic model.Circulation Research, 1990