Analysis of contact mechanics in McKee-Farrar metal-on-metal hip implants
- 1 May 2003
- journal article
- research article
- Published by SAGE Publications in Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
- Vol. 217 (5) , 333-340
- https://doi.org/10.1177/095441190321700501
Abstract
Contact mechanics analysis for a typical McKee-Farrar metal-on-metal hip implant was carried out in this study. The finite element method was used to predict the contact area and the contact pressure distribution at the bearing surfaces. The study investigated the effects of the cement and underlying bone, the geometrical parameters such as the radial clearance between the acetabular cup and the femoral head, and the acetabular cup thickness, as well as other geometrical features on the acetabular cup such as lip and studs. For all the cases considered, the predicted contact pressure distribution was found to be significantly different from that based upon the classical Hertz contact theory, with the maximum value being away from the centre of the contact region. The lip on the cup was found to have a negligible effect on the predicted contact pressure distribution. The presence of the studs on the outside of the cup caused a significant increase in the local contact pressure distribution, and a slight decrease in the contact region. Reasonably good agreement of the predicted contact pressure distribution was found between a three-dimensional anatomical model and a simple two-dimensional axisymmetric model. The interfacial boundary condition between the acetabular cup and the underlying cement, modelled as perfectly fixed or perfectly unbonded, had a negligible effect on the predicted contact parameters. For a given radial clearance of 0.079 mm, the decrease in the thickness of the acetabular cup from 4.5 to 1.5 mm resulted in an increase in the contact half angle from 15° to 26°, and a decrease in the maximum contact pressure from 55 to 20 MPa. For a given acetabular cup thickness of 1.5 mm, a decrease in the radial clearance from 0.158 to 0.0395mm led to an increase in the contact half-angle from 20° to 30°, and a decrease in the maximum contact pressure from 30 to 10 MPa. For zero clearance, although the contact pressure was significantly reduced over most of the contact area, the whole acetabular cup came into contact with the femoral head, leading to stress concentration at the edge of the cup. Design optimization of the geometrical parameters, in terms of the acetabular cup thickness and the radial clearance, is important, not only to minimize the contact stress at the bearing surfaces, but also to avoid equatorial and edge contact.Keywords
This publication has 12 references indexed in Scilit:
- Hip contact forces and gait patterns from routine activitiesPublished by Elsevier ,2001
- Wear and Lubrication of Metal-on-Metal Hip ImplantsClinical Orthopaedics and Related Research, 1999
- Clinical wear behaviour of ultra-high molecular weight polyethylene cups paired with metal and ceramic ball heads in comparison to metal-on-metal pairings of hip joint replacementsProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1997
- Metal-On-Metal Articulation for Artificial Hip Joints: Laboratory Study and Clinical ResultsProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1996
- In Vivo Wear of 3 Types of Metal on Metal Hip Prostheses During 2 Decades of UseClinical Orthopaedics and Related Research, 1996
- Metal on Metal Surface Replacement of the HipClinical Orthopaedics and Related Research, 1996
- Load transfer across the pelvic boneJournal of Biomechanics, 1995
- Contact MechanicsPublished by Cambridge University Press (CUP) ,1985
- An axisymmetric model of acetabular components in total hip arthroplastyJournal of Biomechanics, 1982
- The tribology (friction, lubrication and wear) of all-metal artificial hip jointsWear, 1971