Wear and Lubrication of Metal-on-Metal Hip Implants
- 1 December 1999
- journal article
- research article
- Published by Wolters Kluwer Health in Clinical Orthopaedics and Related Research
- Vol. 369 (369) , 10-24
- https://doi.org/10.1097/00003086-199912000-00003
Abstract
The implication of polyethylene wear particles as the dominant cause of periprosthetic osteolysis has created a resurgence of interest in metal-on-metal implants for total hip arthroplasty because of their potential for improved wear performance. Twenty-two cobalt chromium molybdenum metal-on-metal implants were custom-manufactured and tested in a hip simulator. Accelerated wear occurred within the first million cycles followed by a marked decrease in wear rate to low steady-state values. The volumetric wear at 3 million cycles was very small, ranging from 0.15 to 2.56 mm3 for all implants tested. Larger head-cup clearance and increased surface roughness were associated with increased wear. Independent effects on wear of material processing (wrought, cast) and carbon content were not identified. Implant wear decreased with increasing lambda ratio, a parameter used to relate lubricant film thickness to surface roughness, suggesting some degree of fluid film lubrication during testing. This study provided important insight into the design and engineering parameters that affect the wear behavior of metal-on-metal hip implants and indicated that high quality manufacturing can reproducibly lead to very low wear.Keywords
This publication has 19 references indexed in Scilit:
- Quantitative Assessment of Walking Activity after Total Hip or Knee Replacement*Journal of Bone and Joint Surgery, 1998
- Kinematics of the MATCO™ hip simulator and issues related to wear testing of metal-metal implantsProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1997
- Estimation of Wear in Total Hip Replacement Using a Ten Station Hip SimulatorProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1996
- The Importance of Multidirectional Motion on the Wear of PolyethyleneProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1996
- In Vivo Wear of 3 Types of Metal on Metal Hip Prostheses During 2 Decades of UseClinical Orthopaedics and Related Research, 1996
- 20-Year Results of McKee-Farrar Versus Charnley ProsthesisPublished by Wolters Kluwer Health ,1996
- Metal on Metal Bearings in Hip ArthroplastyClinical Orthopaedics and Related Research, 1996
- Comparison of Alloys and Designs in a Hip Simulator Study of Metal on Metal ImplantsClinical Orthopaedics and Related Research, 1996
- Total hip replacement — past, present and futureBiomaterials, 1982
- A simple theory of asperity contact in elastohydro-dynamic lubricationWear, 1972