A comparison of cardiac output derived from the arterial pressure wave against thermodilution in cardiac surgery patients †
Top Cited Papers
- 1 August 2001
- journal article
- clinical trial
- Published by Elsevier in British Journal of Anaesthesia
- Vol. 87 (2) , 212-222
- https://doi.org/10.1093/bja/87.2.212
Abstract
In three clinical centres, we compared a new method for measuring cardiac output with conventional thermodilution. The new method computes beat-to-beat cardiac output from radial artery pressure by simulating a three-element model of aortic input impedance, and includes non-linear aortic mechanical properties and a self-adapting systemic vascular resistance. We compared cardiac output by continuous model simulation (MF) with thermodilution cardiac output (TD) in 54 patients (18 female, 36 male) undergoing coronary artery bypass surgery. We made three or four conventional thermodilution estimates spread equally over the ventilatory cycle. In 490 series of measurements, thermodilution cardiac output ranged from 2.1 to 9.3, mean 5.0 litre min(-1). MF differed +0.32 (1.0) litre min(-1) on average with limits of agreement of -1.68 and +2.32 litre min(-1). Differences decreased when the first series of measurements in a patient was used to calibrate the model. In 436 remaining series, the mean difference became -0.13 (0.47) litre min(-1) with limits of agreement of -1.05 and +0.79 litre min(-1). When consecutive measurements were made, the change was greater than 0.5 litre min(-1), on 204 occasions. The direction of change was the same with both methods in 199. The difference between the methods remained near zero during surgery suggesting that a single calibration per patient was adequate. Aortic model simulation with radial artery pressure as input reliably monitors changes in cardiac output in cardiac surgery patients. Before calibration, the model cannot replace thermodilution, but after calibration the model method can quantitatively replace further thermodilution estimates.Keywords
This publication has 29 references indexed in Scilit:
- Continuous Cardiac Output CathetersAnesthesiology, 1998
- Continuous cardiac outputCritical Care Medicine, 1998
- Multicenter trial of a new thoracic electrical bioimpedance device for cardiac output estimationCritical Care Medicine, 1994
- Noninvasive Cardiac Output MeasurementAnesthesia & Analgesia, 1992
- An adequate strategy for the thermodilution technique in patients during mechanical ventilationIntensive Care Medicine, 1990
- Frequently repeated fick cardiac output measurements during anesthesiaJournal of Clinical Monitoring and Computing, 1990
- Radial Artery-to-Aorta Pressure Difference after Discontinuation of Cardiopulmonary BypassAnesthesiology, 1989
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- The static elastic properties of 45 human thoracic and 20 abdominal aortas in vitro and the parameters of a new modelJournal of Biomechanics, 1984
- Automatic vs manual injections for thermodilution cardiac output determinationsCritical Care Medicine, 1982