The Humphrey optical coherence tomography scanner: quantitative analysis and reproducibility study of the normal human retinal nerve fibre layer

Abstract
BACKGROUND/AIMS To determine the reproducibility of the Humphrey optical coherence tomography scanner (OCT), software version 5.0, for measurement of retinal nerve fibre layer (RNFL) thickness in normal subjects and to compare OCT measurements with published histological thickness of the human RNFL. METHODS Three independent measurements were obtained at each session for one eye from 15 normal subjects with a mean age of 30.8 (SD 10.9) years. Scans were taken in the peripapillary retina using the default setting (1.74 mm radius from centre of the optic disc) and were repeated 1 week later. Additional scans were obtained at the optic nerve head (ONH) margin overlying the scleral rim, for comparison with available histological data on the human RNFL. RESULTS For the 1.74 mm circular scan, the mean coefficient of variation (COV) for the global RNFL thickness measurement was 5% (SD 3%). This increased to 8% (3%) for quadrant measurements and to 9% (3%) with further subdivision into 12 segments. Significant differences (pCONCLUSION The OCT provides reproducible measurement of the retinal structures that are consistent with the properties of the RNFL. However, comparison with available studies of RNFL thickness in the human suggests that in its present form, the OCT underestimates RNFL thickness. Further refinement of this technology is required to improve the accuracy with which the OCT measures retinal nerve fibre layer thickness.