An elasticity microscope. Part I: Methods
- 1 November 1997
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
- Vol. 44 (6) , 1304-1319
- https://doi.org/10.1109/58.656634
Abstract
An elasticity microscope images tissue stiffness at fine resolution. Possible applications include dermatology, ophthalmology, pathology, and tissue engineering. In addition, if the resolution approaches cellular dimensions, then this system may be very useful in understanding tissue micromorphology. Elasticity images can be reconstructed from displacement and strain fields measured throughout the specimen during controlled external loading. High frequency ultrasound is used to obtain these images by tracking coherent speckle motion during deformation. In this paper, methods are presented to track speckle in two dimensions with near unity correlation coefficients using a high frequency, single element focused transducer. These techniques include improved means for speckle tracking. Procedures to control boundary conditions for consistent specimen deformation and scanning techniques required to obtain a plane-strain state in the imaging plane are also discussed. To test these methods, a 50 MHz elasticity microscope was constructed.Keywords
This publication has 39 references indexed in Scilit:
- Displacement and strain imaging of coronary arteries with intraluminal ultrasoundIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1996
- An axial velocity estimator for ultrasound blood flow imaging, based on a full evaluation of the Doppler equation by means of a two-dimensional autocorrelation approachIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1995
- Adaptive 150 MHz ultrasound imaging of the skin and the eye using an optimal combination of short pulse mode and pulse compression modePublished by Institute of Electrical and Electronics Engineers (IEEE) ,1995
- In vivo imaging of the skin in the 100 MHz region using the synthetic aperture conceptPublished by Institute of Electrical and Electronics Engineers (IEEE) ,1995
- A fundamental limit on the performance of correlation based phase correction and flow estimation techniquesIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1994
- 150 MHz in vivo ultrasound of the skin: imaging techniques and signal processing procedures targeting homogeneous resolutionPublished by Institute of Electrical and Electronics Engineers (IEEE) ,1994
- Quantitative elasticity imagingPublished by Institute of Electrical and Electronics Engineers (IEEE) ,1993
- Speckle tracking in high frequency ultrasound images with application to intravascular imagingPublished by Institute of Electrical and Electronics Engineers (IEEE) ,1993
- The design of protection circuitry for high-frequency ultrasound imaging systemsIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1991
- Acoustic microscope—scanning versionApplied Physics Letters, 1974