Radiation Dose and Image Quality in Pediatric CT: Effect of Technical Factors and Phantom Size and Shape
Top Cited Papers
- 1 November 2004
- journal article
- research article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 233 (2) , 515-522
- https://doi.org/10.1148/radiol.2332032107
Abstract
PURPOSE: To evaluate effects of varying tube current and voltage on radiation dose, image noise, and image contrast with different phantom sizes and shapes. MATERIALS AND METHODS: Four round lucite phantoms with 8–32-cm diameters were scanned with multi–detector row computed tomography (CT) and 80–120 kVp. Radiation dose was based on CT dose index, image noise, and iodine contrast and measured with constant and variable tube currents that were age appropriate for each tube voltage. Radiation dose and image noise and contrast were compared in round and oval 24-cm phantoms. For various combinations of technical factors and phantom sizes and shapes, percentage differences were calculated for radiation dose and image noise and contrast. Associations between tube voltage and radiation dose, image noise, and image contrast in round and oval phantoms were determined by fitting second-degree polynomials to data. Differences in radiation dose and image noise and contrast, which were attributable to differences in tube voltage, were tested with paired t tests. RESULTS: With 165-mAs tube current, radiation doses with 140- and 80-kVp tube voltages were 103% ([41.9 mGy − 20.6 mGy]/20.6 mGy) and 58% ([10.2 mGy − 4.2 mGy]/10.1 mGy) higher in the 8-cm phantom than in the 32-cm phantom. When tube current was adapted for phantom size, radiation dose at 80 kVp in the 8-cm phantom was reduced by 82% ([10.1 mGy − 1.8 mGy]/10.1 mGy). In the 8-cm phantom, tube voltage was decreased from 120 to 80 kVp and tube current remained at 165 mAs, resulting in a 68% noise increase ([3.1 HU − 1.8 HU]/1.8 HU). With variable tube current, 80-kVp tube voltage in the 8-cm phantom led to a 138% noise increase ([7.3 HU − 3.1 HU]/3.1 HU). With reduced tube voltage, image contrast increased. In the 8-cm phantom, with a constant 165-mAs tube current and a decrease in tube voltage from 120 to 80 kVp, there was a 35% ([333 HU − 217 HU]/333 HU) increase in contrast. No difference was noted in radiation dose or noise between round and oval phantoms (P = .604 and P = .06, respectively), but a small statistically significant difference (1%) in contrast attenuation was demonstrated (P = .025). CONCLUSION: Reduced tube voltage for pediatric contrast material–enhanced CT reduces radiation dose and maintains image contrast. Image noise increases, but the effect is minimal in smaller phantoms. An additional reduction in tube current further reduces radiation dose. © RSNA, 2004Keywords
This publication has 28 references indexed in Scilit:
- AAPM/RSNA Physics Tutorial for Residents: Topics in CTRadioGraphics, 2002
- Estimating cancer risks from pediatric CT: going from the qualitative to the quantitativePediatric Radiology, 2002
- Estimates of the cancer risks from pediatric CT radiation are not merely theoretical: Comment on “Point/Counterpoint: In x‐ray computed tomography, technique factors should be selected appropriate to patient size. Against the Proposition” [Med. Phys. 28, 1543–1545 (2001)]Medical Physics, 2001
- Thoracic CT Technique in ChildrenJournal of Thoracic Imaging, 2001
- Radiation Exposures to Patients from CTAmerican Journal of Roentgenology, 2001
- CT scanning: patterns of use and doseJournal of Radiological Protection, 2000
- Low-Dose High-Resolution CT of the Chest in Children and Young AdultsAmerican Journal of Roentgenology, 2000
- Radiation dose reduction in paediatric cranial CTPediatric Radiology, 1999
- Radiation dose reduction in CT of the pediatric pelvis.Radiology, 1994
- Use of radiology in U.S. general short-term hospitals: 1980-1990.Radiology, 1993