Prevention of Hypoglycemia Using Risk Assessment With a Continuous Glucose Monitoring System
- 1 November 2002
- journal article
- Published by American Diabetes Association in Diabetes
- Vol. 51 (11) , 3263-3273
- https://doi.org/10.2337/diabetes.51.11.3263
Abstract
Due to the lag between sugar intake and the beginning of recovery from hypoglycemia, it is necessary to intervene in an anticipatory way if one wants to prevent, not only detect, hypoglycemia. This article presents the principle of a hypoglycemia prevention system based on risk assessment. The risk situation can be defined as the moment when the system estimates that the glucose concentration is expected to reach a hypoglycemia threshold in less than a given time (e.g., 20 min). Since there are well-known discrepancies between blood and interstitial glucose concentrations, the aim of this experimental study performed in nondiabetic rats was first to validate this strategy, and second to determine whether it can work when the glucose concentration is estimated by a glucose sensor in subcutaneous tissue rather than in blood. We used a model of controlled decrease in blood glucose concentration. A glucose infusion, the profile of which mimicked the appearance of glucose from an intragastric load, was administered either when hypoglycemia was detected or on the basis of risk recognition. Despite the lag between the beginning of the load and that of the increase in blood glucose concentration, which was in all experiments 15–20 min, hypoglycemia was fully prevented without overshoot hyperglycemia in the groups of rats in which the glucose load was started when the hypoglycemia risk was detected, on the basis of either blood or interstitial glucose concentration. This was, of course, not the case when the same glucose load was infused at the detection of the hypoglycemia threshold.Keywords
This publication has 17 references indexed in Scilit:
- Do Sensor Glucose Levels Accurately Predict Plasma Glucose Concentrations During Hypoglycemia and Hyperinsulinemia?Diabetes Care, 2002
- Hyperglycemia impairs antro-pyloric coordination and delays gastric emptying in conscious ratsAutonomic Neuroscience, 2002
- Calibration of a subcutaneous amperometric glucose sensor implanted for 7 days in diabetic patientsBiosensors and Bioelectronics, 2001
- Calibration of a subcutaneous amperometric glucose sensorBiosensors and Bioelectronics, 2001
- Recording of subcutaneous glucose dynamics by a viscometric affinity sensorDiabetologia, 2001
- Protracted glucose fall in subcutaneous adipose tissue and skeletal muscle compared with blood during insulin-induced hypoglycaemiaDiabetologia, 1997
- Use of a Subcutaneous Glucose Sensor To Detect Decreases in Glucose Concentration Prior to Observation in BloodAnalytical Chemistry, 1996
- Does fall in tissue glucose precede fall in blood glucose?Diabetologia, 1996
- Dynamic Delay and Maximal Dynamic Error in Continuous BiosensorsAnalytical Chemistry, 1996
- Iatrogenic Hypoglycemia as a Cause of Hypoglycemia-Associated Autonomic Failure in IDDM: A Vicious CycleDiabetes, 1992