Cerebral state monitor, a new small handheld EEG monitor for determining depth of anaesthesia
- 1 March 2006
- journal article
- clinical trial
- Published by Wolters Kluwer Health in European Journal of Anaesthesiology
- Vol. 23 (3) , 208-212
- https://doi.org/10.1017/s0265021505002206
Abstract
The cerebral state index (CSI) derived from a new small handheld electroencephalogram monitor was studied during routine day surgical anaesthesia titrated according to the bispectral index (BIS). The objective was to determine the degree of agreement between the two monitors. Anaesthesia was induced with propofol and fentanyl (0.1 mg) in 38 patients undergoing general anaesthesia for routine day-surgery. Maintenance anaesthesia (sevoflurane (20/38), desflurane (10/38) or propofol (8/38)) titrated by BIS XP (Aspect Medical, Natwick, MA, USA) and BIS and CSI (cerebral State Monitor, Danmeter; Odense, Denmark) index values were recorded every minute. No patient received muscle relaxation. Observer's Assessment of Alertness/Sedation rating scale was used to assess level of sedation. Pair-wise recordings (914) of CSI and BIS were collected. The indices showed similar pattern and decreased with increasing level of sedation, however with large ranges for each level of sedation. Median indices were similar during surgery (BIS: 50 (14-89); CSI: 51 (7-88)) and both indices increased (P 20% from BIS-index in 24% of readings, and on rare occasions CSI indices deviated >100% from the BIS reading. When BIS < 40, CSI decreased slower than BIS and with wider spreading. When used for day-surgery anaesthesia without muscle relaxation, CSI and BIS show similar patterns and numerical values but with the incidence of occasionally large discrepancies between pair-wise readings. Which monitor is the more dependable remains to be established and cannot be implied from this initial explorative study.Keywords
This publication has 22 references indexed in Scilit:
- Cerebral state index during anaesthetic induction: a comparative study with propofol or nitrous oxideActa Anaesthesiologica Scandinavica, 2005
- Comparative Evaluation of the Datex-Ohmeda S/5 Entropy Module and the Bispectral Index® Monitor during Propofol–Remifentanil AnesthesiaAnesthesiology, 2004
- Sevoflurane Decreases Bispectral Index Values More than Does Halothane at Equal MAC MultiplesAnesthesia & Analgesia, 2004
- Spectral Entropy and Bispectral Index as Measures of the Electroencephalographic Effects of SevofluraneAnesthesiology, 2004
- Comparison of BIS and AAI as measures of anaesthetic drug effect during desflurane–remifentanil anaesthesiaActa Anaesthesiologica Scandinavica, 2004
- The Effect of Different Stages of Neuromuscular Block on the Bispectral Index and the Bispectral Index-XP Under Remifentanil/Propofol AnesthesiaAnesthesia & Analgesia, 2004
- Spectral entropy measurement of patient responsiveness during propofol and remifentanil. A comparison with the bispectral index †British Journal of Anaesthesia, 2004
- The Effect of Cerebral Monitoring on Recovery After General Anesthesia: A Comparison of the Auditory Evoked Potential and Bispectral Index Devices with Standard Clinical PracticeAnesthesia & Analgesia, 2003
- Electroencephalographic Bispectral Index Correlates with Intraoperative Recall and Depth of Propofol-Induced SedationAnesthesia & Analgesia, 1997
- Validity and Reliability of the Observer??sJournal of Clinical Psychopharmacology, 1990