Commissioning and quality assurance of the Pinnacle3radiotherapy treatment planning system for external beam photons
- 1 March 2003
- journal article
- research article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 76 (903) , 163-176
- https://doi.org/10.1259/bjr/42085182
Abstract
The commissioning of a Pinnacle3 treatment planning system is described. Four Elekta linear accelerators were commissioned for external beam photons. Measured data were used to derive parameter values for the Pinnacle3 beam model by (1) fitting a Monte Carlo model of the accelerator head to measured data and then extracting the parameters for the Pinnacle3 beam model, and by (2) using the auto-modelling facility within Pinnacle3. Both of these methods yielded dose distributions in accord with published recommendations. A separate small-field beam model, customized for an in-house compact blocking system, was also created, which satisfied appropriate acceptance criteria for stereotactically guided conformal brain treatments. Inhomogeneous, oblique, asymmetrical and irregular fields were also assessed, with calculated and measured doses agreeing to within ±3%. Dose–volume histogram calculation was found to be accurate to within ±5% dose or volume for a grid size of 4 mm × 4 mm × 4 mm, with better accuracy b...Keywords
This publication has 28 references indexed in Scilit:
- 39 oral The use of Monte Carlo generated accelerator data in the photon model of the Pinnacle planning systemRadiotherapy and Oncology, 2001
- The START Trial - measurements in semi-anatomical breast and chest wall phantomsPhysics in Medicine & Biology, 2001
- Quality control of inhomogeneity correction algorithms used in treatment planning systemsInternational Journal of Radiation Oncology*Biology*Physics, 1995
- The development of quality assurance for Sheffield's radiotherapy treatment planning systemsThe British Journal of Radiology, 1993
- Testing a 3-D radiation therapy planning programInternational Journal of Radiation Oncology*Biology*Physics, 1990
- Benchmark measurements for lung dose corrections for x-ray beamsInternational Journal of Radiation Oncology*Biology*Physics, 1988
- Determination of the accuracy of different computer planning systems for treatment with external photon beamsRadiotherapy and Oncology, 1984
- Performance evaluation of computerized treatment planning systems for radiotherapy: external photon beamsInternational Journal of Radiation Oncology*Biology*Physics, 1980
- The storage and manipulation of radiation dose data in a small digital computerThe British Journal of Radiology, 1974