Simultaneous phase correction and SENSE reconstruction for navigated multi‐shot DWI with non‐cartesian k‐space sampling
Open Access
- 7 November 2005
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 54 (6) , 1412-1422
- https://doi.org/10.1002/mrm.20706
Abstract
Phase‐navigated multi‐shot acquisition and parallel imaging are two techniques that have been applied to diffusion‐weighted imaging (DWI) to diminish distortions and to enhance spatial resolution. Specifically, sensitivity encoding (SENSE) has been combined with single‐shot echo planar imaging (EPI). Thus far, it has been difficult to apply parallel imaging methods, like SENSE, to multi‐shot DWI because motion‐induced phase error varies from shot to shot and interferes with sensitivity encoding. Although direct phase subtraction methods have been introduced to correct this phase error, they generally are not suitable for SENSE reconstruction, and they cannot remove all the motion artifacts even if the phase error is fully known. Here, an effective algorithm is proposed to correct the motion‐induced phase error using an iterative reconstruction. In this proposed conjugate‐gradient (CG) algorithm, the phase error is treated as an image encoding function. Given the complex perturbation terms, diffusion‐weighted images can be reconstructed using an augmented sensitivity map. The mathematical formulation and image reconstruction procedures of this algorithm are similar to the SENSE reconstruction. By defining a dynamic composite sensitivity, the CG phase correction method can be conveniently incorporated with SENSE reconstruction for the application of multi‐shot SENSE DWI. Effective phase correction and multi‐shot SENSE DWI (R = 1 to 3) are demonstrated on both simulated and in vivo data acquired with PROPELLER and SNAILS. Magn Reson Med, 2005.Keywords
This publication has 33 references indexed in Scilit:
- Periodically Rotated Overlapping Parallel Lines with Enhanced Reconstruction–Based Diffusion Tensor ImagingJournal of Computer Assisted Tomography, 2004
- Nonlinear phase correction for navigated diffusion imagingMagnetic Resonance in Medicine, 2003
- Generalized autocalibrating partially parallel acquisitions (GRAPPA)Magnetic Resonance in Medicine, 2002
- Advances in sensitivity encoding with arbitrary k‐space trajectoriesMagnetic Resonance in Medicine, 2001
- SENSE: Sensitivity encoding for fast MRIMagnetic Resonance in Medicine, 1999
- Analysis and comparison of motion‐correction techniques in diffusion‐weighted imagingJournal of Magnetic Resonance Imaging, 1996
- Apparent diffusion coefficient mapping of experimental focal cerebral ischemia using diffusion‐weighted echo‐planar imagingMagnetic Resonance in Medicine, 1993
- Echo-Planar Imaging: Magnetic Resonance Imaging in a Fraction of a SecondScience, 1991
- Effect of bulk tissue motion on quantitative perfusion and diffusion magnetic resonance imagingMagnetic Resonance in Medicine, 1991
- The influence of flow and motion in MRI of diffusion using a modified CE‐FAST sequenceMagnetic Resonance in Medicine, 1989