Monte Carlo simulations of a scintillation camera using GATE: validation and application modelling
- 4 September 2003
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 48 (18) , 3021-3042
- https://doi.org/10.1088/0031-9155/48/18/305
Abstract
Geant4 application for tomographic emission (GATE) is a recently developed simulation platform based on Geant4, specifically designed for PET and SPECT studies. In this paper we present validation results of GATE based on the comparison of simulations against experimental data, acquired with a standard SPECT camera. The most important components of the scintillation camera were modelled. The photoelectric effect, Compton and Rayleigh scatter are included in the gamma transport process. Special attention was paid to the processes involved in the collimator: scatter, penetration and lead fluorescence. A LEHR and a MEGP collimator were modelled as closely as possible to their shape and dimensions. In the validation study, we compared the simulated and measured energy spectra of different isotopes: 99mTc, 22Na, 57Co and 67Ga. The sensitivity was evaluated by using sources at varying distances from the detector surface. Scatter component analysis was performed in different energy windows at different distances from the detector and for different attenuation geometries. Spatial resolution was evaluated using a 99mTc source at various distances. Overall results showed very good agreement between the acquisitions and the simulations. The clinical usefulness of GATE depends on its ability to use voxelized datasets. Therefore, a clinical extension was written so that digital patient data can be read in by the simulator as a source distribution or as an attenuating geometry. Following this validation we modelled two additional camera designs: the Beacon transmission device for attenuation correction and the Solstice scanner prototype with a rotating collimator. For the first setup a scatter analysis was performed and for the latter design, the simulated sensitivity results were compared against theoretical predictions. Both case studies demonstrated the flexibility and accuracy of GATE and exemplified its potential benefits in protocol optimization and in system design.Keywords
This publication has 10 references indexed in Scilit:
- Combination of MCNP and SimSET for Monte Carlo simulation of SPECT with medium- and high-energy photonsIEEE Transactions on Nuclear Science, 2002
- Asymmetric cone-beam transmission tomographyIEEE Transactions on Nuclear Science, 2001
- Relevance of accurate Monte Carlo modeling in nuclear medical imagingMedical Physics, 1999
- An object-oriented Monte Carlo simulator for 3D cylindrical positron tomographsComputer Methods and Programs in Biomedicine, 1999
- ROOT — An object oriented data analysis frameworkNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1997
- Physically realistic Monte Carlo simulation of source, collimator and tomographic data acquisition for emission computed tomographyPhysics in Medicine & Biology, 1992
- PETSIM: Monte Carlo simulation of all sensitivity and resolution parameters of cylindrical positron imaging systemsPhysics in Medicine & Biology, 1992
- Monte Carlo techniques in medical radiation physicsPhysics in Medicine & Biology, 1991
- A Monte Carlo program for the simulation of scintillation camera characteristicsComputer Methods and Programs in Biomedicine, 1989
- Gamma-Ray Detection Efficiency and Image Resolution in Sodium IodideReview of Scientific Instruments, 1964