Evaluation of a Candidate Reference Measurement Procedure for Serum Free Testosterone Based on Ultrafiltration and Isotope Dilution–Gas Chromatography–Mass Spectrometry
Open Access
- 1 November 2004
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 50 (11) , 2101-2110
- https://doi.org/10.1373/clinchem.2004.037358
Abstract
Background: To assess the analytical validity of free testosterone (FTe) measurements, a reference measurement procedure (RMP) is required. For steroids, isotope dilution–mass spectrometry is accepted as state-of-the-art technology. Because FTe is defined as the hormone fraction in serum water in equilibrium with the protein-bound fraction, the RMP should include a physical separation step. The use of equilibrium dialysis (ED) or ultrafiltration (UF) is advocated. Our objective was to develop such a candidate RMP. Methods: We selected UF combined with isotope dilution–gas chromatography–mass spectrometry (ID-GC/MS) for direct measurement of Te in the ultrafiltrate. After optimization of the UF process, the complete procedure was validated by use of split-sample comparisons with indirect ED (iED) and symmetric dialysis (SyD). Results: The candidate RMP gave maximum within-day, between-day, and total CVs of 3.0%, 3.1%, and 4.3%. The Deming regression equations for the respective method comparisons were: UF-ID-GC/MS = 0.98(iED) − 53 pmol/L (r = 0.94; Sy|x= 42 pmol/L) and UF-ID-GC/MS = 0.92(SyD) + 21 pmol/L (r = 0.97; Sy|x= 31 pmol/L). Conclusions: We achieved the objective of a state-of-the-art candidate RMP, which agreed well with iED and SyD. However, we also demonstrated that a degree of discordance remains, which may require a decision from an authoritative organization on the recommended procedure to measure free hormone concentrations.Keywords
This publication has 25 references indexed in Scilit:
- Calculation of free and bound fractions of testosterone and estradiol-17β to human plasma proteins at body temperaturePublished by Elsevier ,2003
- A Critical Evaluation of Simple Methods for the Estimation of Free Testosterone in SerumJournal of Clinical Endocrinology & Metabolism, 1999
- Errors in the Measurement of Plasma Free Testosterone 1Journal of Clinical Endocrinology & Metabolism, 1997
- Measurement of plasma free steroids by direct radioimmunoassay of ultrafiltrate in association with the monitoring of free components with [14C]glucoseClinica Chimica Acta; International Journal of Clinical Chemistry, 1991
- Measurement of Free Hormones in BloodEndocrine Reviews, 1990
- The Free Hormone Hypothesis: A Physiologically Based Mathematical Model*Endocrine Reviews, 1989
- A theoretically sound and practicable equilibrium dialysis method for measuring percentage of free testosteroneThe Journal of Steroid Biochemistry and Molecular Biology, 1985
- A simplified method for the calculation of unbound or free testosterone by equilibrium dialysis of undiluted plasmaClinica Chimica Acta; International Journal of Clinical Chemistry, 1985
- The application of rate dialysis to the determination of free steroids in plasmaAnalytical Biochemistry, 1983
- SEX‐HORMONE‐BINDING GLOBULINClinical Endocrinology, 1974