Nondispersive Isotope-Selective Infrared Spectroscopy: A New Analytical Method for13C-Urea Breath Tests
- 1 January 1996
- journal article
- research article
- Published by Taylor & Francis in Scandinavian Journal of Gastroenterology
- Vol. 31 (5) , 442-445
- https://doi.org/10.3109/00365529609006762
Abstract
Background: Currently, stable isotope techniques in breath tests using 13C-labeled substrates are limited to a few centers equipped with expensive and complex isotope ratio mass spectrometry (IRMS). Although breath samples can be mailed to these centers, widespread application of 13C breath tests would be more feasible with a cheaper and more practicable analysis system at hand. Methods: We therefore tested the newly developed nondispersive isotope-selective infrared spectrometer (NDIRS) with reference to IRMS in a clinical setting comparing the results of both techniques in 538 consecutive 13C-urea breath tests performed for the detection of Helicobacter pylori infection. Results: With NDIRS five false-positive and three false-negative results were observed; that is, the sensitivity of NDIRS was 98.3%, and the specificity was 98.6%. The delta over base-line values of both devices correlated linearly (Y = 0.87 ± 0.01 X + 0.29 ± 0.15; r = 0.95; p < 0.0001; n = 538). Conclusions: When running this large number of breath tests in 3 days, the NDIRS proved to be a reliable, stable, and easy-to-operate analytical tool, which is well qualified for gastroenterologic application in the diagnostic routine. Both the price and the easy handling of NDIRS will facilitate the widespread use of the noninvasive stable isotope technique for 13C breath tests.Keywords
This publication has 13 references indexed in Scilit:
- Isotope-selective non-dispersive infrared spectrometry for detection of Helicobacter pylori infection with 13C-urea breath testThe Lancet, 1995
- The [13C]acetate breath test accurately reflects gastric emptying of liquids in both liquid and semisolid test mealsGastroenterology, 1995
- Biomedical Application of an Isotope Selective Nondispersive Infrared Spectrometer for 13CO2 and 12CO2 Concentration Measurements in Breath SamplesIsotopenpraxis Isotopes in Environmental and Health Studies, 1994
- A Sensitive Isotope Selective Nondispersive Infrared Spectrometer for 13CO2 and 12CO2 Concentration Measurements in Breath SamplesIsotopenpraxis Isotopes in Environmental and Health Studies, 1994
- Pharmacological modulation of gastric emptying rate of solids as measured by the carbon labelled octanoic acid breath test: influence of erythromycin and propantheline.Gut, 1994
- Laser-Based Analysis of Carbon Isotope RatiosScience, 1994
- Mixed triglyceride breath test: A noninvasive test of pancreatic lipase activity in the duodenumGastroenterology, 1989
- A Carbon-13 Breath Test to Characterize Glucose Absorption and Utilization in ChildrenJournal of Pediatric Gastroenterology and Nutrition, 1988
- Measurement of carbon-13 dioxide/carbon-12 dioxide abundance by nondispersive infrared heterodyne ratiometry as an alternative to gas isotope ratio mass spectrometryAnalytical Chemistry, 1986
- Caffeine: A Model Compound for Measuring Liver FunctionHepatology, 1984