Comparative evaluation of the Abbott TDX, the Abbott ABA200, and the Syva LAB5000 for assay of serum gentamicin
- 1 January 1987
- journal article
- research article
- Published by Oxford University Press (OUP) in Journal of Antimicrobial Chemotherapy
- Vol. 19 (1) , 127-133
- https://doi.org/10.1093/jac/19.1.127
Abstract
Three automated chemical assays for serum gentamicin were compared for accuracy, reproducibility and cost. One method utilized fluorescence polarization (Abbott TDX), and the other two enzyme-multiplied immunoassay (Abbott ABA200, and the Syva Lab5000). All three systems produced a high degree of accuracy and reproducibility with spiked samples when the concentrations of gentamicin were within the range of 3.0–8.0 mg/l. However, with concentrations below 2.0 mg/l or above 8.0 mg/l, only the TDX system gave acceptable coefficients of variation and accurate recoveries. Similarly, excellent correlations were obtained between all three systems for assays of clinical specimens containing 2.0–8.0 mg/l gentamicin, but above and below this range, the correlations were poor except between TDX and Lab5000 within the range of 0.0–2.0 mg/l. The Abbott TDX was thus the most accurate and reproducible of the three systems for the assay of serum gentamicin in the critical concentrations below 2.0 mg/l and above 8.0 mg/l. The cost per assay by the Abbott TDX was US$4.57 compared with US$5.40 for the Abbott ABA200, and US$3.20 for the Syva Lab5000.Keywords
This publication has 3 references indexed in Scilit:
- Comparative analysis of two rapid, automated methods for determining aminoglycoside levelsJournal of Clinical Microbiology, 1984
- Comparison of enzyme-multiplied immunoassay technique with fluorescence polarization immunoassay for determination of gentamicin and tobramycin levels in serumJournal of Clinical Microbiology, 1984
- Evaluation of four gentamicin and tobramycin assay procedures for clinical laboratoriesJournal of Clinical Microbiology, 1983