Automated assessment of blood pressure using BpTRU compared with assessments by a trained technician and a clinic nurse
- 1 October 2005
- journal article
- research article
- Published by Wolters Kluwer Health in Blood Pressure Monitoring
- Vol. 10 (5) , 257-262
- https://doi.org/10.1097/01.mbp.0000173486.44648.b2
Abstract
To determine the accuracy and reproducibility of a new automated blood pressure manometer (BpTRU) relative to auscultatory blood pressure assessed by a research nurse and to that assessed by a clinic nurse. Firefighters in a cohort study had blood pressure assessed on up to five occasions with BpTRU and by a trained research technician. Patients in an internal medicine clinic had blood pressure assessed by the clinic nurse and by BpTRU. The absolute values of blood pressure, reproducibility and effect on hypertension classification were compared with the different methods. The research technician readings were higher than the BpTRU readings at visit 1 (3.0/2.7 mmHg, P<0.0001) but the readings converged by visits 4–5 because of a greater reduction in the research nurse readings. The BpTRU readings had similar reproducibility and classification of hypertension as the research technician but did not exhibit terminal digit preference while the research technician readings did. The BpTRU had substantially lower readings (8/7 mmHg) and fewer hypertensive readings than those of the nurse in the internal medicine clinic. This preliminary study found that the BpTRU had desirable characteristics that suggest that it would be a suitable replacement for auscultatory assessment of blood pressure in clinical practice. A large confirmatory study performed in a usual clinic setting is required.Keywords
This publication has 13 references indexed in Scilit:
- Does evidence-based medicine suggest that physicians should not be measuring blood pressure in the hypertensive patient?American Journal of Hypertension, 2004
- Clinical decision-making in hypertension using an automated (BpTRU™) measurement deviceJournal of Human Hypertension, 2003
- Use of an automated blood pressurerecording device, the BpTRU, to reduce the“white coat effect” in routine practiceAmerican Journal of Hypertension, 2003
- The Seventh Report of the Joint National Committee on Prevention, Detection, Evaluation, and Treatment of High Blood PressureThe JNC 7 ReportJAMA, 2003
- Comparison of the oscillometric blood pressure monitor (BPM-100 Beta ) with the auscultatory mercury sphygmomanometerBlood Pressure Monitoring, 2001
- Validation of a new algorithm for the BPM-100 electronic oscillometric office blood pressure monitorBlood Pressure Monitoring, 2001
- Comparison of nurse- and physician-determined clinic blood pressure levels in patients referred to a hypertension clinicJournal Of Hypertension, 2000
- Is usual measurement of blood pressure meaningful?Blood Pressure Monitoring, 1999
- Relationship between blood pressure, as measured by a nurse, by a doctor and under basal conditions, and micro-albuminuria and left ventricular massJournal Of Hypertension, 1991
- EFFECTS OF BLOOD-PRESSURE MEASUREMENT BY THE DOCTOR ON PATIENT'S BLOOD PRESSURE AND HEART RATEThe Lancet, 1983