A phased array coil for human cardiac imaging
- 1 July 1995
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 34 (1) , 92-98
- https://doi.org/10.1002/mrm.1910340114
Abstract
A prototype cardiac phased array receiver coil was constructed that comprised a cylindrical array and a separate planar array. Both arrays had two coil loops with the same coil dimensions. Data acquisition with the cylindrical array placed on the human chest, and the planar array placed under the back, yielded an overall enhancement of the signal‐to‐noise ratio (SNR) over the entire heart by a factor of 1.1–2.85 over a commercially available flexible coil and a commercially available four‐loop planar phased array coil. This improvement in SNR can be exploited in cardiac imaging to increase the spatial resolution and reduce the image acquisition time.Keywords
This publication has 13 references indexed in Scilit:
- Cardiac Tagging with Breath‐Hold Cine MRIMagnetic Resonance in Medicine, 1992
- An NMR Phased Array for Human Cardiac 31P SpectroscopyMagnetic Resonance in Medicine, 1992
- Volume imaging with MR phased arraysMagnetic Resonance in Medicine, 1991
- The NMR phased arrayMagnetic Resonance in Medicine, 1990
- Phosphate metabolite imaging and concentration measurements in human heart by nuclear magnetic resonanceMagnetic Resonance in Medicine, 1990
- Human heart: tagging with MR imaging--a method for noninvasive assessment of myocardial motion.Radiology, 1988
- Switched surface coil system for bilateral MR imaging.Radiology, 1988
- Phase and sensitivity of receiver coils in magnetic resonance imagingMedical Physics, 1986
- Measurement of signal intensities in the presence of noise in MR imagesMedical Physics, 1985
- A signal-to-noise calibration procedure for NMR imaging systemsMedical Physics, 1984