Evaluation of COBAS AmpliPrep nucleic acid extraction in conjunction with COBAS AmpliScreen HBV DNA, HCV RNA and HIV-1 RNA amplification and detection
- 1 November 2005
- journal article
- research article
- Published by Wiley in Vox Sanguinis
- Vol. 89 (4) , 193-200
- https://doi.org/10.1111/j.1423-0410.2005.00691.x
Abstract
This report describes the evaluation of the COBAS AmpliPrep instrument for fully automated generic nucleic acid extraction in conjunction with hepatitis B virus (HBV) DNA, hepatitis C virus (HCV) RNA, and human immunodeficiency virus (HIV)-1 RNA COBAS AmpliScreen amplification and detection using serial dilutions of the WHO international standards (IS) and the PeliCheck reference panels.Serial diluted samples of the WHO IS and the PeliCheck reference panels were tested 24 times to determine the HBV DNA, HCV RNA and HIV-1 RNA detection limits by Probit analysis. The existence and extent of cross-contamination were assessed by testing alternating high titre HBV DNA-positive and -negative samples. The specificity of the AmpliPrep-AmpliScreen test for HBV was determined by testing 232 minipools consisting of six donations, all negative for HCV/HIV-1 nucleic acid testing (NAT) and HBsAg. In addition, a HBV genotypes A-G panel was tested.The respective 95% detection limits (and 95% CI) on the WHO IS and on the PeliCheck reference panels were 6.7 (4.3-13) IU/ml and 123 (68-301) gEq/ml for HBV DNA, 23 (11-106) IU/ml and 126 (84-233) gEq/ml for HCV RNA, and 187 (108-422) IU/ml and 183 (108-434) gEq/ml for HIV-1 RNA. Based on the WHO IS and the PeliCheck reference panels, no significant differences in sensitivity for HBV and HCV were found between AmpliPrep and the licensed MultiPrep extraction method. The sensitivity of AmpliPrep-AmpliScreen for HIV-1 was probably twofold lower as compared to the MultiPrep-AmpliScreen method. No cross contamination was observed. All 232 minipools were HBV NAT-negative. The AmpliPrep-AmpliScreen test for HBV detected HBV genotypes A-G with equal sensitivity.The AmpliPrep instrument combined with the AmpliScreen assays for HBV, HCV and HIV-1 is robust and suitable for NAT donor screening. The sensitivity criteria for HIV-1 and HCV as defined by the Paul Ehrlich Institute and the Food and Drug Administration for minipool NAT screening are met by this system. SINGLE SENTENCE SUMMARY: Generic COBAS AmpliPrep nucleic acid extraction in conjunction with COBAS AmpliScreen detection for HBV, HCV and HIV-1.Keywords
This publication has 21 references indexed in Scilit:
- Multicenter performance evaluation of a transcription‐mediated amplification assay for screening of human immunodeficiency virus‐1 RNA, hepatitis C virus RNA, and hepatitis B virus DNA in blood donationsTransfusion, 2005
- International Forum: 1Vox Sanguinis, 2005
- Lack of correlation between HBsAg and HBV DNA levels in blood donors who test positive for HBsAg and anti‐HBc: implications for future HBV screening policyTransfusion, 2004
- Diversity and origin of hepatitis B virus in Dutch blood donorsJournal of Medical Virology, 2004
- Quantification of Parvovirus B19 DNA Using COBAS AmpliPrep Automated Sample Preparation and LightCycler Real-Time PCRThe Journal of Molecular Diagnostics, 2004
- Should HBV DNA NAT replace HBsAg and/or anti-HBc screening of blood donors?Transfusion Clinique et Biologique, 2004
- Evaluation of an Automated Sample Preparation Protocol for Quantitative Detection of Hepatitis C Virus RNAJournal of Clinical Microbiology, 2002
- Implementation of donor screening for infectious agents transmitted by blood by nucleic acid technologyVox Sanguinis, 2002
- An international collaborative study to establish a World Health Organization international standard for hepatitis B virus DNA nucleic acid amplification techniquesVox Sanguinis, 2001