Clinical Performance of Steroid-Eluting Pacing Leads with 1.2-mm2 Electrodes
- 1 November 1997
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 20 (11) , 2799-2809
- https://doi.org/10.1111/j.1540-8159.1997.tb05438.x
Abstract
To raise pacing impedance and reduce battery current drain, new tined steroid-eluting leads were developed with 1.2-mm2 hemispherical electrodes, instead of conventional 5-8 mm2. Twenty-two unipolar J-shaped atrial leads and 25 unipolar ventricular leads (models 4533 and 4033, respectively) were implanted in 33 consecutive patients and followed for a mean of 25 months (range 18-29). Handling characteristics of atrial leads were found favorable. The leads slipped easily into the right atrial appendage and were easy to position. Handling characteristics of ventricular leads were satisfying, but more efforts had to be applied to cross the tricuspid valve. Special care was taken to avoid perforation of the myocardium due to the small lead tip. Following implantation, four ventricular and one atrial lead exhibited instability of pacing thresholds that resolved spontaneously within 1-3 days of implantation. Except for this, no lead malfunctioned. The reoperation rate was zero. The mean electrogram amplitudes of 15 mV (ventricle) and 4 mV (atrium), and the mean chronic pacing threshold of 0.085 ms at 1.6 V (app. 0.43 V at 0.5 ms) were comparable with the best values seen in the literature on passive fixation leads. The rest of the electrophysiological parameters were enhanced: mean pacing impedances were 984 omega (acute) and 900 Q (chronic), mean slew rates 3.26 V/s (ventricle) and 1.75 V/s (atrium), mean acute voltage threshold at 0.5 ms was 0.25 V, mean current and energy thresholds calculated at 0.5 ms were 260 microA and 32 nJ (acute) and 478 microA and 103 nJ (chronic). The electrical characteristics of these leads provide for increased pacemaker longevity in combination with substantial safety margins for pacing and sensing.Keywords
This publication has 20 references indexed in Scilit:
- Initial Experience with 1.5-mm2 High Impedance, Steroid-luting Pacing ElectrodesPacing and Clinical Electrophysiology, 1996
- POSTER SESSION VPacing and Clinical Electrophysiology, 1994
- Atrial and Ventricular Stimulation Threshold Development: A Comparative Study in Patients with a DDD Pacemaker and Two Identical Carbon‐Tip LeadsPacing and Clinical Electrophysiology, 1990
- Stimulation Characteristics of a Steroid-Eluting Electrode Compared with Three Conventional ElectrodesPacing and Clinical Electrophysiology, 1990
- Pacing and Sensing: How Can One Electrode Fulfill Both Requirements?Pacing and Clinical Electrophysiology, 1987
- Clinical Experience with 910 Carbon Tip Leads: Comparison with Polished Platinum Leads.Pacing and Clinical Electrophysiology, 1986
- Optimal Pacemaker Sensing with Respect to Amplitude and Slew Rate of Intracardiac Electrograms: Theoretical Analysis by Computer SimulationPacing and Clinical Electrophysiology, 1984
- Electrode‐Heart Interface: Definition of the Ideal ElectrodePacing and Clinical Electrophysiology, 1983
- The Interdependence Between Electrogram, Total Electrode Impedance and Pacemaker Input Impedance Necessary to Obtain Adequate Functioning of Demand PacemakersPacing and Clinical Electrophysiology, 1979
- CLINICAL USE OF A NEW TRANSVENOUS ELECTRODE*Annals of the New York Academy of Sciences, 1969