Evaluation of Plasma Resistant Hollow Fiber Membranes For Artificial Lungs
- 1 September 2004
- journal article
- research article
- Published by Wolters Kluwer Health in Asaio Journal
- Vol. 50 (5) , 491-497
- https://doi.org/10.1097/01.mat.0000138078.04558.fe
Abstract
Hollow fiber membranes (HFMs) used in artificial lungs (oxygenators) undergo plasma leakage (or wetting) in which blood plasma slowly fills the pores of the fiber wall, plasma leaks into gas pathways, and overall gas exchange decreases. To overcome this problem plasma resistant fibers are being developed that are skinned asymmetric or composite symmetric versions of microporous oxygenator fibers. This report evaluates several candidate plasma resistant HFMs in terms of their gas permeance and plasma resistance as measured in a surfactant wet out test. Five candidate fibers were compared with each other and with a control fiber. CO2 and O2 gas permeance (in ml/s/cm2/cm Hg) in the plasma resistant fibers ranged from 3.15E-04 to 1.71E-03 and 3.40E-04 to 1.08E-03, respectively, compared with 1.62E-02 and 1.77E-02 for the control fiber. Maximum dye bleed through for the plasma resistant fibers in the forced wet out test were significantly less than for the control fiber. CO2 gas permeance of a plasma resistant fiber imposes the greatest constraint upon artificial lung design for sufficient gas exchange. However, our results suggest sufficient plasma resistance can be achieved using special skinned and composite HFMs while maintaining an acceptable CO2 gas permeance for a broad range of artificial lung applications.Keywords
This publication has 18 references indexed in Scilit:
- A respiratory gas exchange catheter: In vitro and in vivo tests in large animalsThe Journal of Thoracic and Cardiovascular Surgery, 2002
- Early Experience with a Polymethyl Pentene Oxygenator for Adult Extracorporeal Life SupportAsaio Journal, 2002
- In Vivo Uptake and Elimination of Isoflurane by Different Membrane Oxygenators during Cardiopulmonary BypassAnesthesiology, 2002
- High risk of intraoperative awareness during cardiopulmonary bypass with isoflurane administration via diffusion membrane oxygenatorsPerfusion, 2002
- Testing of an Intrathoracic Artificial Lung in a Pig ModelPublished by Wolters Kluwer Health ,1996
- Experimental Evaluation of a Newly Developed Ultrathin Silicone Layer Coated Hollow Fiber OxygenatorAsaio Journal, 1996
- Gas permeability of hollow fiber membranes in a gas-liquid systemJournal of Membrane Science, 1996
- Effects of Blood Phase Oscillation on Gas Transfer in a Microporous Intravascular LungAsaio Journal, 1994
- Plasma Leakage through Microporous MembranesAsaio Journal, 1992
- Technical Aspects of Plasma Leakage Prevention in Microporous Capillary Membrane OxygenatorsAnnual Northeast Bioengineering Conference, 1989