High risk of intraoperative awareness during cardiopulmonary bypass with isoflurane administration via diffusion membrane oxygenators
- 1 May 2002
- journal article
- research article
- Published by SAGE Publications in Perfusion
- Vol. 17 (3) , 175-178
- https://doi.org/10.1191/0267659102pf566oa
Abstract
In cardiac surgery with the aid of extracorporeal circulation (ECC), inhalation anaesthetics can be administered via the oxygenator. Until the recent advent of a new type of diffusion membrane oxygenator, we routinely added the inhalation agent, isoflurane, to the gas flow of a micro-porous capillary membrane-type oxygenator. Applying this procedure to the diffusion-type oxygenators, the depth of anaesthesia appeared to be affected, which manifested itself through unusually high intraoperative perfusion pressures. This observation led to a prospective randomized study comprising 60 patients and two models of a microporous capillary membrane oxygenator, as well as two models of a diffusion membrane oxygenator. Simultaneous isoflurane concentration measurements at both the gas inlet and outlet ports of the oxygenators showed that, whereas in the microporous capillary-type oxygenators the isoflurane administered was reduced by about 50% during the passage of gas through the device, there was only a minimal transfer of isoflurane in the diffusion-type membrane oxygenators.Keywords
This publication has 11 references indexed in Scilit:
- Desflurane pharmacokinetics during cardiopulmonary bypassJournal of Cardiothoracic and Vascular Anesthesia, 2001
- Evidence for Preconditioning by Isoflurane in Coronary Artery Bypass Graft SurgeryCirculation, 1999
- Effects of Rapid Increases of Desflurane and Sevoflurane to Concentrations of 1.5 MAC on Systemic Vascular Resistance and Catecholamine Response during Cardiopulmonary BypassAnesthesiology, 1997
- Effects of destflurane and isoflurane on systemic vascular resistance during hypothermic cardiopulmonary bypassJournal of Cardiothoracic and Vascular Anesthesia, 1997
- In vitro uptake and elimination of isoflurane by different membrane oxygenatorsJournal of Cardiothoracic and Vascular Anesthesia, 1996
- Small Pore Size Microporous Membrane Oxygenator Reduces Plasma Leakage during Prolonged Extracorporeal Circulation: A Case ReportThe International Journal of Artificial Organs, 1996
- Effects of sevoflurane and isoflurane on systemic vascular resistance: use of cardiopulmonary bypass as a study modelBritish Journal of Anaesthesia, 1996
- Isoflurane Uptake and Elimination Are Delayed by Absorption of Anesthetic by the Scimed Membrane OxygenatorAnesthesia & Analgesia, 1989
- Technical Aspects of Plasma Leakage Prevention in Microporous Capillary Membrane OxygenatorsAnnual Northeast Bioengineering Conference, 1989
- Washin and washout of isoflurane during cardiopulmonary bypassCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1988