Glycolysis inhibitors negatively bias blood glucose measurements: potential impact on the reported prevalence of diabetes mellitus
- 27 June 2007
- journal article
- research article
- Published by BMJ in Journal of Clinical Pathology
- Vol. 60 (7) , 820-923
- https://doi.org/10.1136/jcp.2006.039925
Abstract
Background:Blood glucose concentrations are essential in defining diabetes mellitus. Recent guidelines advocate either of two discrete methods for sample collection and processing. One of these involves addition of glycolysis inhibitors, such as sodium fluoride–potassium oxalate (NaF–KOx) to sample collection tubes, whereas the other requires immediate refrigeration and sample separation.Aims:To examine whether the choice of the preanalytical process has any impact on subsequent glucose determinations.Methods:62 healthy men participated in the study during screening for diabetes. Paired venous blood samples were collected in a serum-gel tube and a tube containing NaF–KOx (both Sarstedt, Leicester, UK). Serum was promptly separated from gel tube samples and refrigerated, whereas NaF–KOx samples were not separated until immediately before analysis. Glucose concentrations were determined using an Olympus AU 2700 analyser incorporating an automated hexokinase method.Results:Mean (95% CI) glucose concentration in serum-gel tube samples was 5.2 mmol/l (5.0 to 5.4 mmol/l), whereas the concentration in tubes containing NaF–KOx was 4.9 mmol/l (4.8 to 5.1 mmol/l). A negative bias of 0.23 mmol/l (0.16 to 0.30 mmol/l) and relative negative bias of 4.7 % (3.2% to 6.3%) were observed for samples collected in NaF–KOx tubes, consistent with the combined effects of glycolysis and dilution.Conclusions:Bias associated with the use of NaF–KOx tubes may have a significant impact on the prevalence of fasting hyperglycaemia, according to current diagnostic criteria. The small but significant difference between preanalytical processes should be considered when screening for the presence of diabetes mellitus.Keywords
This publication has 16 references indexed in Scilit:
- Follow-up Report on the Diagnosis of Diabetes MellitusDiabetes Care, 2003
- Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes MellitusClinical Chemistry, 2002
- Optimization of preanalytical conditions and analysis of plasma glucose. 1. Impact of the new WHO and ADA recommendations on diagnosis of diabetes mellitusScandinavian Journal of Clinical and Laboratory Investigation, 2001
- Effect of Specimen Collection and Storage on Blood Glucose and Lactate Concentrations in Healthy, Hyperthyroid and Diabetic CatsVeterinary Clinical Pathology, 2000
- Graphical interpretation of analytical data from comparison of a field method with a Reference Method by use of difference plotsClinical Chemistry, 1997
- IMPROVEMENT OF GLUCOSE PRESERVATION IN BLOOD-SAMPLES1990
- Stability of Plasma Glucose during StorageAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 1990
- HANDLING OF BLOOD SPECIMENS FOR GLUCOSE ANALYSIS1990
- Effectiveness of sodium fluoride as a preservative of glucose in blood.Clinical Chemistry, 1989
- Manual of American Society of Clinical Pathologists Workshop on GlucoseAmerican Journal of Clinical Pathology, 1956