Validation in Volunteers of a Near-Infrared Spectroscope for Monitoring Brain Oxygenation In Vivo
- 1 February 1996
- journal article
- neurosurgical anesthesia
- Published by Wolters Kluwer Health in Anesthesia & Analgesia
- Vol. 82 (2) , 269-277
- https://doi.org/10.1097/00000539-199602000-00010
Abstract
Cerebral oximeters based on near-infrared spectroscopy may provide a continuous, noninvasive assessment of cerebral oxygenation. We evaluated a prototype cerebral oximeter (Invos 3100; Somanetics, Troy, MI) in 22 conscious, healthy volunteers breathing hypoxic gas mixtures. Using the first 12 subjects (training group), we developed an algorithm based on the mathematic relationship that converts detected light from the field surveyed by the probe to cerebral hemoglobin oxygen saturation (CSf O2). To develop the algorithm, we correlated the oximeter result with the estimated combined brain hemoglobin oxygen saturation (CScomb O2, where CScomb O2 = Sa O2 times 0.25 + Cj O2 times 0.75 and Sj O2 = jugular venous saturation). We then validated the algorithm in the remaining 10 volunteers (validation group). A close association (r2 = 0.798-0.987 for individuals in the training group and r2 = 0.794-0.992 for individuals in the validation group) existed between CSf O2 and CScomb O2. We conclude that continuous monitoring with cerebral oximetry may accurately recognize decreasing cerebral hemoglobin oxygen saturation produced by systemic hypoxemia. (Anesth Analg 1996;82:269-77)Keywords
This publication has 23 references indexed in Scilit:
- THE INFLUENCE OF POSITION CHANGE OF NEAR-INFRARED SPECTROSCOPIC ASSESSMENT OF CEREBRAL HEMOGLOBIN DESATURATIONAnesthesiology, 1994
- Near-Infrared Spectroscopy of Adult Brains Requires Wide Optode SeparationAnesthesiology, 1994
- Near infrared spectroscopy in adultsAnaesthesia, 1993
- Regional cerebrovascular oxygen saturation measured by optical spectroscopy in humans.Stroke, 1991
- Noninvasive cerebral optical spectroscopy for monitoring cerebral oxygen delivery and hemodynamicsCritical Care Medicine, 1991
- Influence of path length on remote optical sensing of properties of biological tissueApplied Optics, 1989
- Epidural HydromorphoneAnesthesiology, 1987
- Cerebral monitoring in newborn infants by magnetic resonance and near infrared spectroscopyScandinavian Journal of Clinical and Laboratory Investigation, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Noninvasive, Infrared Monitoring of Cerebral and Myocardial Oxygen Sufficiency and Circulatory ParametersScience, 1977